首页> 外文期刊>Urology >Prospective evaluation of refurbished flexible ureteroscope durability seen in a large public tertiary care center with multiple surgeons
【24h】

Prospective evaluation of refurbished flexible ureteroscope durability seen in a large public tertiary care center with multiple surgeons

机译:前瞻性评估在大型公共三级医疗中心中由多名医生进行翻新的柔性输尿管镜的耐用性

获取原文
获取原文并翻译 | 示例
       

摘要

Objective To evaluate the durability and cost of maintenance for outsourced, refurbished flexible ureteroscopes. Materials and Methods Ureteroscope usage and repair were prospectively recorded over a 365-day period at a large 836-bed public hospital. Cases were performed by 14 different urologists using either refurbished DUR-8 or DUR-8 Elite model ureteroscopes. Retrograde cases involving calculi, urothelial carcinoma, stricture, and diagnostic evaluations were included. Ureteroscope repairs were performed by a single outsourced repair vendor, not the original manufacturer. Results A total of 501 ureteroscopic cases involving 550 ureteroscope usages were performed over a 365-day period. Semirigid ureteroscopes were used for 281 (56.1%) cases and refurbished flexible ureteroscopes for 220 (43.9%). The reason for the ureteroscopy was calculi in 386 (77.0%) cases, urothelial carcinoma in 32 (6.4%), stricture in 36 (7.2%), and diagnostic in 47 (9.4%). No repairs were needed during this period for semirigid scopes. Ureteral access sheaths were used in 82 (37.7%) of the cases. A total of 32 instances of catastrophic breakage occurred. Each newly refurbished ureteroscope was used for an average of 6.9 times before incurring further damage requiring repair. Conclusion Refurbished flexible ureteroscopes that have undergone comprehensive repair are extremely fragile in the setting of multiple surgeon users in a large public hospital that uses central processing for sterilization and storage. This poor durability results in significant maintenance, repair, and administrative inconvenience that should be considered along with the purchase price.
机译:目的评估外包,翻新的柔性输尿管镜的耐用性和维护成本。材料和方法前瞻性地记录了一家大型836张病床的公立医院在365天的时间内对输尿管镜的使用和维修情况。由14位不同的泌尿科医师使用翻新的DUR-8或DUR-8 Elite型输尿管镜进行手术。包括结石,尿路上皮癌,狭窄和诊断评估的逆行病例。输尿管镜维修由单个外包维修供应商进行,而不是原始制造商。结果在365天的时间内,共进行了501例输尿管镜检查,其中550例使用了输尿管镜。半刚性输尿管镜用于281例(56.1%),翻新的柔性输尿管镜用于220例(43.9%)。输尿管镜检查的原因是结石386例(77.0%),尿路上皮癌32例(6.4%),狭窄36例(7.2%),诊断47例(9.4%)。在此期间,对于半刚性示波器,无需维修。 82例(37.7%)患者使用输尿管进入鞘。总共发生了32次灾难性破坏。每个新近翻新的输尿管镜平均使用6.9次,然后再造成需要维修的损坏。结论在使用中央处理进行消毒和存储的大型公立医院中,经过多位外科医生使用的情况下,经过全面修复的翻新的柔性输尿管镜非常脆弱。这种较差的耐用性会导致大量的维护,修理和管理上的不便,应与购买价格一起考虑。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号