Keratometry evaluations with the Pentacam high resolution in comparison with the automated keratometry and conventional corneal topography.

机译:与自动角膜测定法和常规角膜地形图相比,使用Pentacam高分辨率进行角膜测定法评估。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To determine the reliability and repeatability of keratometry (K) measurements obtained with the Pentacam high resolution (HR), automated keratometry, and corneal topography systems. METHODS: The right eyes of 46 healthy subjects were examined prospectively. Keratometry measurements in the flat (Kf) and steep (Ks) meridians were taken by 2 independent investigators with the Pentacam HR (Oculus, Wetzlar, Germany) followed by automated keratorefractometry (KR-8100; Topcon, Tokyo, Japan), and corneal topography (TMS-4; Tomey, Erlangen, Germany). RESULTS: The mean K readings of the Pentacam HR, automated keratometry, and corneal topography were 43.40/43.34 diopter (D), 43.99/43.98 D, and 43.80/43.83 D, respectively. The difference between the values was statistically significant (P < 0.0001, repeated measures analysis of variance). Strong significant correlation was observed between the Pentacam HR and keratometry (Kf: r = 0.952/0.954; Ks: r = 0.845, Spearman rank test), and Pentacam HR and corneal topography (Kf: r = 0.933/0.930; Ks: r = 0.838/0.829) (P < 0.0001). No significant difference was presented between the 2 investigators for any of the instruments (P = 0.215-0.983). Moreover, high correlation was found between the K readings of the observers (interoperator intraclass correlation coefficients ranged from 0.95 to 0.99). CONCLUSIONS: The Pentacam HR provided reliable K measurements in clinical practice in comparison with an automated keratometer and a corneal topographer. Based on the results, for patient follow-up, one keratometry device is recommended.
机译:目的:确定使用Pentacam高分辨率(HR),自动角膜测量和角膜地形图系统获得的角膜测量(K)测量结果的可靠性和可重复性。方法:前瞻性检查46名健康受试者的右眼。两名独立研究人员使用Pentacam HR(Oculus,Wetzlar,德国)在平坦(Kf)和陡峭(Ks)子午线上进行角膜测量,然后进行自动角膜折光法(KR-8100; Topcon,日本东京)和角膜地形图(TMS-4;德国埃尔兰根的Tomey)。结果:Pentacam HR,自动角膜曲率法和角膜地形图的平均K读数分别为43.40 / 43.34屈光度(D),43.99 / 43.98 D和43.80 / 43.83D。值之间的差异具有统计学意义(P <0.0001,重复测量方差分析)。观察到Pentacam HR与角膜曲率法之间存在强烈的显着相关性(Kf:r = 0.952 / 0.954; Ks:r = 0.845,Spearman等级检验),以及Pentacam HR与角膜地形图(Kf:r = 0.933 / 0.930; Ks:r = 0.838 / 0.829)(P <0.0001)。 2名研究者之间对于任何仪器均无显着差异(P = 0.215-0.983)。此外,在观察者的K读数之间发现高度相关(互操作员组内相关系数的范围为0.95至0.99)。结论:与自动角膜曲率计和角膜地形图仪相比,Pentacam HR在临床实践中提供了可靠的K测量。根据结果​​,建议对患者进行随访,以便进行患者随访。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号