Effect of fluorescein dye staining of the tear film on Scheimpflug measurements of central corneal thickness.

机译:泪膜荧光素染料染色对中央角膜厚度Scheimpflug测量的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To assess whether fluorescein dye in the tear film results in a clinically relevant change in central corneal thickness (CCT), as measured with Scheimpflug imaging (Pentacam HR; Oculus, Wetzlar, Germany). The study was conducted at Moorfields Eye Hospital NHS Foundation Trust. METHODS: In this prospective, randomized, controlled, single-center study on healthy volunteers, CCT was measured using rotating Scheimpflug imaging and partial coherence interferometry technique. After 3 baseline measurements of both eyes, a drop of 0.25% fluorescein with 0.5% proxymetacaine hydrochloride dye, as used for measuring intraocular pressure with applanation Goldmann tonometry, was instilled in the first eye and measurements were repeated after 1 minute and 5, 10, 20, 40, and 60 minutes. A fluorescein drop was then instilled in the second eye, and 1 measurement was performed after 1 minute. Immediately afterwards, the second eye was rinsed with 2 ml of balanced saline solution and the CCT measurement was repeated. On a different day, a Schirmer II test was performed. RESULTS: Thirty eyes of 15 volunteers were included in this study. The average CCT measured with Scheimpflug imaging was 539.1 mum (SD, 32.2 mum), and after instilling fluorescein, CCT values increased by 46.6 mum (SD, 11.4 mum; P < 0.01, paired t test). In 10 of 15 subjects, the fluorescein layer was still present 40 minutes after instilling the drop. Rinsing the eye decreased the fluorescein layer by 49.6% and a value of 24.8 mum (SD, 16.8 mum). CONCLUSIONS: Because CCT measurements are critical for corneal refractive surgery, Scheimpflug imaging should not be performed after use of fluorescein dye for staining of the corneal epithelium or for applanation tonometry.
机译:目的:评估泪膜中的荧光素染料是否导致中央角膜厚度(CCT)的临床相关变化,如Scheimpflug成像(Pentacam HR; Oculus,Wetzlar,德国)所测量的。该研究是在Moorfields眼科医院NHS基金会信托基金进行的。方法:在这项针对健康志愿者的前瞻性,随机,对照,单中心研究中,使用旋转Scheimpflug成像和部分相干干涉测量技术对CCT进行了测量。在对两只眼睛进行3次基线测量后,在第一只眼睛中滴入一滴0.25%的荧光素和0.5%的代卡他卡因盐酸盐,用于压平Goldmann眼压法测量眼内压,并在1分钟和5、10, 20、40和60分钟。然后将荧光素滴剂滴入第二只眼,并在1分钟后进行1次测量。之后,立即用2 ml平衡盐溶液冲洗第二只眼睛,并重复进行CCT测量。在另一天,进行了Schirmer II测试。结果:15名志愿者的三十只眼被纳入这项研究。用Scheimpflug成像测得的平均CCT为539.1微米(SD,32.2微米),滴注荧光素后,CCT值增加了46.6微米(SD,11.4微米; P <0.01,配对t检验)。在15名受试者中的10名中,滴注液滴40分钟后,荧光素层仍然存在。冲洗眼睛可使荧光素层减少49.6%,值为24.8微米(标准差,16.8微米)。结论:由于CCT测量对于角膜屈光手术至关重要,因此在使用荧光染料染色角膜上皮或压平眼压计后不应进行Scheimpflug成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号