Corneal endothelium evaluation with 2 noncontact specular microscopes and their semiautomated methods of analysis.

机译:用2台非接触式镜面显微镜评估角膜内皮及其半自动化分析方法。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: Noncontact specular microscopes and semiautomated methods of endothelial analysis are widely used in the clinical practice and for research purposes. In this study, the interexaminer reproducibility between two noncontact specular microscopes using different semiautomated methods was assessed in normal corneas. METHODS: Average cell size (ACS), endothelial cell density (ECD), coefficient of variation (CV), and hexagonality values were independently calculated by two examiners with the Topcon SP 2000P Image-NET retraced method and the Konan CC7000 center method in 49 normal subjects. Interexaminer reproducibility and correlation between instruments/methods were assessed through 95% limits of agreement, intra-class correlation coefficient (ICC), Pearson correlation coefficient, and Spearman rank test. RESULTS: Interexaminer reproducibility was good for ACS and ECD measurements (ICC > 0.85) but weak for CV and hexagonality (ICC < 0.65). Significantly higher ACS and lower ECD values (P < 0.05) were obtained with the Topcon in comparison with the Konan method by both examiners. ACS and ECD mean differences were respectively, +26 microm2 and -184 cell/mm2 for examiner 1 and +36 microm2 and -228 cell/mm2 for examiner 2. No significant difference (P > 0.05) between instruments/methods was found in CV or hexagonality values; however, the correlation for these parameters was low (r < 0.20). CONCLUSIONS: Either instrument or method allows reproducible and accurate ACS and ECD measurements on normal corneas, but the difference between the two instruments is systematic and significant. The difference found were not clinically meaningful however for research/longitudinal study purposes, the data collected with the two systems cannot be used interchangeably.
机译:目的:非接触式镜面显微镜和半自动化的内皮分析方法广泛用于临床实践和研究目的。在这项研究中,在正常角膜中使用不同的半自动化方法评估了两个非接触式镜面显微镜之间的检查者间的可重复性。方法:由两名检查员分别使用Topcon SP 2000P Image-NET回溯法和Konan CC7000中心法在49中独立计算平均细胞大小(ACS),内皮细胞密度(ECD),变异系数(CV)和六边形值。正常人。通过95%的一致性限制,组内相关系数(ICC),Pearson相关系数和Spearman等级检验,评估了检查者间的可重复性和仪器/方法之间的相关性。结果:检查者间的可重复性对于ACS和ECD测量良好(ICC> 0.85),但对于CV和六角形却很弱(ICC <0.65)。与两名检查员的Konan方法相比,Topcon显着提高了ACS,降低了ECD值(P <0.05)。 ACS和ECD的平均差异分别为,检验员1为+26 microm2和-184 cell / mm2,检验员2为+36 microm2和-228 cell / mm2,在CV中没有发现仪器/方法之间的显着差异(P> 0.05)。或六边形值;但是,这些参数的相关性很低(r <0.20)。结论:两种仪器或方法均可对正常角膜进行可重复且准确的ACS和ECD测量,但两种仪器之间的差异是系统性的且意义重大。发现的差异在临床上没有意义,但是出于研究/纵向研究的目的,两个系统收集的数据不能互换使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号