Central corneal thickness measurements: using an ultrasonic instrument and 4 optical instruments.

机译:中央角膜厚度测量:使用超声波仪器和4台光学仪器。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Both between-session and within-session repeatability were determined for measurement of corneal thickness with the following techniques: ultrasound pachymetry (UP), confocal microscopy (CS), Orbscan pachymetry (ORB), spectral oscillation interferometry (SOI), and optical coherence tomography (OCT). METHODS: The right eyes of 20 normal subjects were tested on 2 different days. For each session, 2 central corneal thickness measurements were determined using the ConfoScan 3 microscope, the Orbscan system, and the Stratus OCT system; 5 measurements were found with the Sonogage ultrasound pachymeter; and between 2 and 10 measurements were found with the SOI system. RESULTS: This study showed that SOI had the best repeatability, whereas CS had the worst repeatability for both within-session and between-session repeatability. Based on 95% limits of agreement (LoA), the within-session repeatability can be ranked as: (1) SOI; (2) UP; (3) OCT; (4) Orbscan pachymetry; and (5) CS. SOI had a bias of 0.13 and 95% LoA of -1.07 to 1.33, whereas UP had a bias of -0.98 and 95% LoA of -7.35 to 5.40. The between-session repeatability of the instruments can be ranked as: (1) SOI; (2) OCT; (3) UP; (4) Orbscan pachymetry; and (5) CS. Of the optical methods, SOI had the best repeatability with a bias of 1.26 and 95% LoA of -6.14 to 8.66, followed by OCT with a bias of 0.75 and 95% LoA of -16.22 to 17.27. CONCLUSIONS: Of the 4 instruments that are commercially available, the UP was the most repeatable for within sessions, and the OCT was the most repeatable for between sessions.
机译:目的:使用以下技术确定会话间和会话内重复性,以测量角膜厚度:超声测厚法(UP),共聚焦显微镜(CS),Orbscan测厚法(ORB),光谱振荡干涉法(SOI)和光学相干断层扫描(OCT)。方法:在2天的不同时间对20名正常受试者的右眼进行了测试。对于每个疗程,使用ConfoScan 3显微镜,Orbscan系统和Stratus OCT系统确定2个中央角膜厚度测量值。 Sonogage超声波测厚仪发现5次测量;使用SOI系统进行2到10次测量。结果:这项研究表明,SOI具有最佳的可重复性,而CS的会话内和会话间可重复性均最差。根据95%的协议限制(LoA),会话内的可重复性可以归类为:(1)SOI; (2)向上; (3)华侨城; (4)Orbscan测厚法; (5)CS。 SOI具有0.13的偏差和-1.07至1.33的95%LoA,而UP具有-0.98的偏差和-7.35至5.40的95%LoA。仪器的会话间可重复性可归类为:(1)SOI; (2)华侨城; (3)向上; (4)Orbscan测厚法; (5)CS。在光学方法中,SOI具有最佳的可重复性,偏差为1.26,95%LoA为-6.14至8.66,其次是OCT,偏差为0.75,95%LoA为-16.22至17.27。结论:在市售的4种仪器中,UP在会议期间的重复性最高,OCT在会议之间的重复性最高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号