首页> 外文期刊>Optics Letters >Pupil-segmentation-based adaptive optical correction of a high-numerical-aperture gradient refractive index lens for two-photon fluorescence endoscopy
【24h】

Pupil-segmentation-based adaptive optical correction of a high-numerical-aperture gradient refractive index lens for two-photon fluorescence endoscopy

机译:基于光瞳分割的双光子荧光内窥镜高数值孔径渐变折射率透镜的自适应光学校正

获取原文
获取原文并翻译 | 示例
       

摘要

The intrinsic aberrations of high-NA gradient refractive index (GRIN) lenses limit their image quality as well as field of view. Here we used a pupil-segmentation-based adaptive optical approach to correct the inherent aberrations in a two-photon fluorescence endoscope utilizing a 0.8 NA GRIN lens. By correcting the field-dependent aberrations, we recovered diffraction-limited performance across a large imaging field. The consequent improvements in imaging signal and resolution allowed us to detect fine structures that were otherwise invisible inside mouse brain slices.
机译:高NA梯度折射率(GRIN)透镜的固有像差会限制其图像质量以及视野。在这里,我们使用基于光瞳分割的自适应光学方法来校正使用0.8 NA GRIN透镜的双光子荧光内窥镜中的固有像差。通过校正与场有关的像差,我们在大成像场上恢复了衍射受限的性能。随之而来的成像信号和分辨率的改善使我们能够检测出在老鼠脑切片内不可见的精细结构。

著录项

  • 来源
    《Optics Letters》 |2012年第11期|共3页
  • 作者

    Chen Wang; Na Ji;

  • 作者单位
  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-19 13:46:47

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号