首页> 外文期刊>Critical care medicine >Point-of-care glucose measurement systems should be used with great caution in critically ill intensive care unit patients.
【24h】

Point-of-care glucose measurement systems should be used with great caution in critically ill intensive care unit patients.

机译:危重症重症监护病房患者应谨慎使用即时护理葡萄糖测量系统。

获取原文
获取原文并翻译 | 示例
           

摘要

Meynaar et al assessed the accuracy of the AccuChek point-of-care glucose measurement system in the intensive care unit (ICU) setting (1). The authors conclude that the AccuChek glucose measurement in whole blood with a correction factor has acceptable accuracy for use in ICU patients. We recently demonstrated that the AccuChek (using plasma referenced test strips) failed the International Organization for Standardization (ISO) criteria in our study in >13% of samples (2). The ICU population that was studied by Meynaar et al had lower Acute Physiology and Chronic Health Evaluation II scores on hospital admission compared to our study (17.8 +- 6.7 vs. 21.0 [16-25]) and included a higher number of surgical patients (47% vs. 16%). Because patients with inaccurate glucose values had a higher disease severity score and mortality, the discrepancy between the results found by Meynaar and our study group may be related to the less severely ill study population. This is supported by the fact that in the same study the AccuChek was highly accurate in non-ICU patients (2.5% inaccurate samples according to ISO standards). Because we found different accuracy levels for the same test device between different populations, it is seems likely that the confounding factor is patient related.
机译:Meynaar等人评估了重症监护病房(ICU)设置中AccuChek即时血糖测量系统的准确性(1)。作者得出的结论是,使用校正因子对全血进行AccuChek葡萄糖测量对于ICU患者而言具有可接受的准确性。我们最近证明,在我们的研究中,AccuChek(使用以血浆为参考的测试条)未通过国际标准化组织(ISO)的标准,超过了13%的样品(2)。与我们的研究相比,Meynaar等人研究的ICU人群入院时的急性生理和慢性健康评估II评分较低(17.8 +-6.7与21.0 [16-25]),并且手术患者的数量更高( 47%和16%)。由于血糖值不准确的患者的疾病严重程度评分和死亡率较高,因此Meynaar与我们的研究小组发现的结果之间的差异可能与病情较轻的研究人群有关。在同一项研究中,AccuChek在非ICU患者中非常准确(根据ISO标准,不准确的样本为2.5%),这证明了这一点。因为我们在不同人群之间针对同一测试设备发现了不同的准确性水平,所以混淆因素似乎与患者有关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号