A method has been developed to improve the accuracy with which the polarization state of light can be characterized by the rotating quarter-wave plate technique. Through detailed analysis, verified by experiment, we determine the positions of the optic axes of the retarder and linear polarizer, and the wave plate retardance, to better than 1 deg for typical signal-to-noise ratios. Accurate determination of the Stokes parameters can be achieved using a single wave plate for a wide range of optical wavelengths using this technique to determine the precise retardance at each of the wavelengths of interest.
机译:已经开发出一种方法来提高精度,通过旋转四分之一波片技术可以表征光的偏振态。通过详细的分析,并通过实验验证,我们确定了延迟器和线性偏振器的光轴位置以及波片延迟,对于典型的信噪比,该延迟要优于1度。斯托克斯参数的准确确定可以使用单个波片来实现,该波板可以在很宽的光波长范围内使用该技术来确定每个感兴趣波长处的精确延迟。