Uniform guidelines for the data processing and validation of qualitative and quantitative multi-residue analysis using full-spectrum high-resolution mass spectrometry are scarce. Through systematic research, optimal mass accuracy and sensitivity are obtained after refining the postprocessing of the HRMS data. For qualitative analysis, transforming the raw profile spectra to centroid spectra is recommended resulting in a 2.3 fold improved precision on the accurate mass determination of spectrum peaks. However, processing centroid data for quantitative purposes could lead to signal interruption when too narrow mass windows are applied for the construction of extracted ion chro-matograms. Therefore, peak integration on the raw profile data is recommended. An optimal width of the mass window of 50 ppm, which is a trade-off between sensitivity and selectivity, was obtained for a TOF instrument providing a resolving power of 20,000 at full width at half maximum (FWHM). For the validation of HRMS analytical methods, widespread concepts such as the signal-to-noise ratios for the determination of decision limits and detection capabilities have shown to be not always applicable because in some cases almost no noise can be detected anymore. A statistical methodology providing a reliable alternative is extended and applied.
机译:缺乏使用全光谱高分辨率质谱法进行数据处理以及定性和定量多残留分析验证的统一指南。通过系统研究,在优化HRMS数据的后处理之后,可以获得最佳的质量准确性和灵敏度。对于定性分析,建议将原始谱图谱转换为质心谱图,从而在准确测定质谱峰的质量上提高2.3倍。但是,如果将太窄的质量窗口应用于提取离子色谱图的构建,则出于定量目的处理质心数据可能会导致信号中断。因此,建议对原始配置文件数据进行峰积分。对于TOF仪器,获得的质量窗口的最佳宽度为50 ppm,这是灵敏度和选择性之间的折衷。TOF仪器在半峰全宽(FWHM)下提供20,000的分辨能力。为了验证HRMS分析方法,已显示出广泛的概念,例如用于确定决策极限和检测能力的信噪比,并不总是适用的,因为在某些情况下几乎不再能检测到噪声。提供可靠替代方案的统计方法得到扩展和应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号