A single pump column-switching technique coupled with polystyrene-divinylbenzene-carbon nanotubes column for the determination of trace anions in different concentrated organic matrices by ion chromatography

机译:单泵色谱柱切换技术结合聚苯乙烯-二乙烯基苯-碳纳米管色谱柱,用于离子色谱法测定不同浓缩有机基质中的痕量阴离子

获取原文
获取原文并翻译 | 示例
       

摘要

An ion chromatographic method with on-line sample pretreatment was developed for the trace analysis of seven common anions in concentrated matrices. The pretreatment column used in this study consisted of polystyrene-divinylbenzene (PS-DVB) and multi-walled carbon nanotubes (MWCNTs). It was too hydrophobic to retain different inorganic anions, but it showed a strong affinity for organic compounds. Thus, this chromatographic system could be used to detect trace anions in organic solvents, organic acids and relevant salts. The addition of MWCNTs decreased the surface areas of stationary phases and the retention times of organic matrices were shortened. Compared with conventional column-switching technique, only a single instrument (1CS2100) was needed in this system, including a pump, a conductivity detector, an eluent generator, a six-port valve and a ten-port valve. An electrochemical self-generating suppressor (ESGS) was adopted to convert the eluent of KOH into water for the matrix elimination. Two different eluent were employed in the chromatographic system, one for separation and the other for matrix elimination. The sample pretreatment and analysis were realized simultaneously. After optimization of this system, a calibration study was conducted by preparing and analyzing eight concentrations (between 5 and 5000 μg L~(-1)) of mixture standards of seven anions in deionized water. The linearity was between 0.9990 and 0.9998, and the detection limits ranged from 0.41 to 3.17 μg L~(-1). A spiking study was performed on three representative organic chemicals with satisfactory recoveries between 88.1% and 118.5% when the concentrations of the matrices did not exceed 10g L~(-1).
机译:开发了一种在线样品预处理的离子色谱方法,用于痕量分析浓缩基质中的七个常见阴离子。本研究中使用的预处理塔由聚苯乙烯-二乙烯基苯(PS-DVB)和多壁碳纳米管(MWCNT)组成。它的疏水性太强,无法保留不同的无机阴离子,但对有机化合物表现出很强的亲和力。因此,该色谱系统可用于检测有机溶剂,有机酸和相关盐类中的痕量阴离子。 MWCNT的添加减少了固定相的表面积,并缩短了有机基质的保留时间。与传统的列切换技术相比,该系统仅需要一个仪器(1CS2100),包括泵,电导率检测器,洗脱液发生器,六通阀和十通阀。采用电化学自生抑制剂(ESGS)将KOH的洗脱液转化为水以消除基质。色谱系统中使用了两种不同的洗脱液,一种用于分离,另一种用于基质消除。同时实现了样品的预处理和分析。优化该系统后,通过制备和分析去离子水中七种阴离子的八种浓度的混合标准物(在5和5000μgL〜(-1)之间)进行校准研究。线性在0.9990至0.9998之间,检出限在0.41至3.17μgL〜(-1)之间。当基质的浓度不超过10g L〜(-1)时,对三种代表性的有机化学品进行了加标研究,回收率在88.1%至118.5%之间,令人满意。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号