An improved procedure for flow-based turbidimetric sulphate determination based on a liquid core waveguide and pulsed flows

机译:基于液芯波导和脉冲流的基于流量的浊度硫酸盐测定的改进程序

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

An improved flow-based procedure is proposed for turbidimetric sulphate determination in waters.The flow system was designed with solenoid micro-pumps in order to improve mixing conditions and minimize reagent consumption as well as waste generation.Stable baselines were observed in view of the pulsed flow characteristic of the systems designed with solenoid micro-pumps,thus making the use of washing solutions unnecessary.The nucleation process was improved by stopping the flow prior to the measurement,thus avoiding the need of sulphate addition.When a 1-cm optical path flow cell was employed,linear response was achieved within 20-200mgL~(-1),described by the equation S=-0.0767+0.00438C (mg L~(-1)),r=0.999.The detection limit was estimated as 3 mg L~(-1) at the 99.7% confidence level and the coefficient of variation was 2.4% (n=20).The sampling rate was estimated as 33 determinations per hour.A long pathlength (100-cm) flow cell based on a liquid core waveguide was exploited to increase sensitivity in turbidimetry.Baseline drifts were avoided by a periodical washing step with EDTA in alkaline medium.Linear response was observed within 7-16mgL~(-1),described by the equation S =-0.865+0.132C (mgL~(-1)),r=0.999.The detection limit was estimated as 150 mu g L~(-1) at the 99.7% confidence level and the coefficient of variation was 3.0% (n = 20).The sampling rate was estimated as 25 determinations per hour.The results obtained for freshwater and rain water samples were in agreement with those achieved by batch turbidimetry at the 95% confidence level.
机译:提出了一种改进的基于流动的方法来测定水中的浊度硫酸盐。该流动系统设计了螺线管微型泵,以改善混合条件并最大程度地减少试剂消耗和废物产生,并在脉冲下观察到稳定的基线设计有螺线管微型泵的系统的流量特性,因此无需使用洗涤液。通过在测量之前停止流量来改善成核过程,从而避免了添加硫酸盐的要求。当光程为1 cm时采用流通池,线性响应在20-200mgL〜(-1)范围内,由方程S = -0.0767 + 0.00438C(mg L〜(-1))描述,r = 0.999。置信水平为99.7%时为3 mg L〜(-1),变异系数为2.4%(n = 20)。估计采样率为每小时33次测定基于长光程(​​100-cm)流通池在液芯波导管上被利用来增加比浊法的灵敏度。通过在碱性介质中使用EDTA进行定期洗涤步骤避免了基线漂移。在7-16mgL〜(-1)范围内观察到线性响应,由方程S = -0.865 + 0.132C(mgL〜(- 1)),r = 0.999。在99.7%置信水平下,检出限估计为150μg L〜(-1),变异系数为3.0%(n = 20)。采样率估计为25每小时测定的淡水和雨水样品的结果与分批比浊法在95%置信度下获得的结果一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号