Naked-Eye Coulometric Sensor Using a Longitudinally Oriented Ag Band Electrode in a Microfluidic Channel

机译:在微流控通道中使用纵向取向的Ag带电极的裸眼库仑传感器

获取原文
获取原文并翻译 | 示例
       

摘要

In this Article, we report a coulometric sensing platform that is capable of sensing analytes on a working electrode and providing a visual readout of the analyte concentration on a silver (Ag) band counter electrode in a microchannel. The display mechanism relies on the electro-oxidation of metallic Ag as a complementary reaction to the sensing reduction reaction. The Ag band counter electrode is arranged longitudinally in a microchannel while the frontal tip of the band electrode directly faces a gold (Au) working electrode, which lies across the microchannel. The Ag oxidation always occurs at the band electrode's tip region that faces the working electrode due to the Ohmic potential drop across the solution in the microchannel. The decrement of the Ag electrode, which is dearly measurable with the naked eye, correlates linearly with an analyte concentration (e.g., 0.1-2.5 mM p-benzoquinone) and with an analyte feeding rate (i.e., a sample solution flow rate of 1.0-75.0 mu L min(-1)). The platform design is also effective for a model analyte of horseradish peroxidase (HRP)-avidin in the dynamic range of 0.1-3.0 mu g mL(-1).
机译:在本文中,我们报告了一种库仑感测平台,该平台能够感应工作电极上的分析物,并在微通道中的银(Ag)谱带对电极上提供分析物浓度的可视读数。显示机制依赖于金属Ag的电氧化,作为感应还原反应的补充反应。 Ag带对电极纵向排列在微通道中,而带电极的前端直接面对横跨微通道的金(Au)工作电极。由于微通道中溶液上的欧姆电势下降,Ag氧化总是在面对工作电极的带状电极尖端区域发生。用肉眼几乎可以测量的Ag电极的减量与分析物浓度(例如0.1-2.5 mM对苯醌)和分析物进料速度(即样品溶液流速为1.0- 75.0μL min(-1))。该平台设计对于动态范围为0.1-3.0μg mL(-1)的辣根过氧化物酶(HRP)-亲和素模型分析物也有效。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号