The detection system which enables simultaneous fluorescence polarization (FP) measurement of multiple samples was proposed and proven by a proof-of-concept experiment on the viscosity dependence of FP of fluorescein sample in water-ethylene glycol solution and another experiment on the FP immunoassay of prostaglandin E2 sample. The measurement principle of FP is based on the synchronization between the orientation of the liquid crystal molecules and the sampling frequency of a CCD. This report is the first description of the simultaneous FP measurement of multiple samples. This system has a great potential for equipment miniaturization and price reduction as well as providing simultaneous FP measurement of multiple samples.
机译:提出了一种能够同时测量多个样品的荧光偏振(FP)的检测系统,并通过关于荧光素样品FP在水-乙二醇溶液中的粘度依赖性的概念验证实验以及另一个针对FP免疫分析的实验进行了验证。前列腺素E2样品。 FP的测量原理是基于液晶分子的取向与CCD的采样频率之间的同步。该报告是对多个样品同时进行FP测量的首次描述。该系统在设备小型化和价格降低以及同时对多个样品进行FP测量方面具有巨大的潜力。