The development of spectroscopic and electrochemical devices that can accommodate very small samples is of considerable importance to many areas of science and technology. We report here on the design and characteristics of a simple apparatus for the electrochemical and spectroelectrochemical analysis of submicro-liter aqueous samples. The device is easily assembled from common laboratory materials and equipment, utilizing a bifurcated fiber-optic probe, standard disk electrodes of millimeter dimensions, and a polymer electrolyte film salt bridge to enable the analysis of nanoliter-scale sample volumes in a thin-layer configuration. Excellent performance has been demonstrated via measurements on aqueous ferricyanide solutions using sample volumes as low as 20 nL.
机译:可以容纳非常小的样品的光谱和电化学设备的开发对许多科学和技术领域都具有重要意义。我们在这里报告了用于亚微升水样的电化学和光谱电化学分析的简单设备的设计和特性。该设备易于使用普通的实验室材料和设备进行组装,利用分叉的光纤探头,毫米级的标准圆盘电极和聚合物电解质膜盐桥实现了薄层配置中纳升级样品体积的分析。通过使用低至20 nL的样品量对铁氰化物水溶液进行测量,已证明了优异的性能。