Detection of Trace Proteins in Multicomponent Mixtures Using Displacement Chromatography

机译:置换色谱法检测多组分混合物中的痕量蛋白质

获取原文
获取原文并翻译 | 示例
       

摘要

Model protein feed mixtures containing three abundant and seven trace proteins at various concentrations were identified and employed in a series of displacement experiments. Reversed-phase liquid chromatography (RPLC) and matrix assisted laser desorption ionization-time-of-flight (MALDI-TOF) mass spectrometry were used to evaluate the compositions of both the feed mixtures and effluent fractions from the displacement experiments. The results demonstrated that trace proteins were focused at the boundaries between the abundant solutes where they were enriched and concentrated. For many of the multicomponent feed mixtures, mass spectrometry analyses of the displacement column effluent fractions resulted in the identification of trace proteins that were not detectable in the feed. In addition, the use of minimal or no salt in the carrier solutions enabled the analysis of displacement fractions by direct infusion mass spectrometry. These results are significant in that they indicate that while the presence of abundant proteins can often be problematic for the detection of trace components, displacement chromatography may be able to employ these abundant proteins to focus trace proteins in the displacement train, thus facilitating detection.
机译:鉴定了包含三种丰富和七个痕量蛋白质(不同浓度)的模型蛋白质进料混合物,并将其用于一系列置换实验。反相液相色谱(RPLC)和基​​质辅助激光解吸电离飞行时间(MALDI-TOF)质谱用于评估置换实验中进料混合物和流出物馏分的组成。结果表明,痕量蛋白质集中于富集和浓缩的丰富溶质之间的边界。对于许多多组分进料混合物,置换柱流出物馏分的质谱分析导致了进料中无法检测到的痕量蛋白质的鉴定。另外,在载体溶液中使用最少的盐或不使用盐就可以通过直接注入质谱法分析置换分数。这些结果是有意义的,因为它们表明尽管丰富的蛋白质的存在对于痕量组分的检测通常是有问题的,但置换色谱法可能能够利用这些丰富的蛋白质将痕量蛋白质聚焦在置换序列中,从而促进检测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号