首页> 外文期刊>Analytical chemistry >Certification of Beryllium Mass Fraction in SRM 1877 Beryllium Oxide Powder Using High-Performance Inductively Coupled Plasma Optical Emission Spectrometry with Exact Matching
【24h】

Certification of Beryllium Mass Fraction in SRM 1877 Beryllium Oxide Powder Using High-Performance Inductively Coupled Plasma Optical Emission Spectrometry with Exact Matching

机译:使用高性能电感耦合等离子体发射光谱与精确匹配技术,对SRM 1877氧化铍粉末中的铍质量分数进行认证

获取原文
获取原文并翻译 | 示例
       

摘要

High-performance inductively coupled plasma optical emission spectrometry (HP-ICP-OES) was used to certify the Be mass fraction in National Institute of Standards and Technology (NIST) Standard Reference Material (SRM) 1877 Beryllium Oxide Powder. The certified value and expanded uncertainty expressed at a 95percent confidence level is (0.3576 +- 0.0024) g/g. To obtain best results, the Be mass fractions, Mn (internal standard) mass fractions, and matrix compositions of the calibration solutions were carefully matched to those of the sample solutions for each individual HP-ICP-OES analysis. This "exact matching" approach was used because experience at NIST has shown that it often affords improved accuracy and precision in HP-ICP-OES analysis. NIST has never published these observations. Due to the toxicity of BeO and the difficulty of containing the very fine powder material, sets of solutions for HP-ICP-OES analysis were prepared by laboratories collaborating with NIST who have the experience and equipment needed to work with the material safely. Each laboratory utilized a unique digestion protocol(s). After preparing the sets of solutions, the collaborating laboratories shipped them to NIST for HP-ICP-OES analysis. NIST provided the collaborating laboratories with solution preparation kits and spreadsheets to help establish traceability of the HP-ICP-OES results to the International System of Units (SI) and to allow exact matching to be accomplished. The agreement observed among the four individual Be mass fraction values determined from the sets of solutions prepared by the collaborating laboratories was 0.074percent relative (1s of mean). The excellent agreement provides a measure of confidence in the robustness of each of the digestion procedures, as well as in the certified Be mass fraction value. The analytical benefits of using exact matching for this particular certification were investigated. Results show that exactly matching the matrix compositions of the standards to the samples for each HP-ICP-OES analysis was critical to obtaining the excellent agreement observed among the individual Be mass fraction values and also helped to minimize bias and uncertainty in the certified value. Unlike previous NIST studies, exactly matching the Be and Mn mass fractions of the standards to the samples for this particular certification appears to have had little effect on the data.
机译:高性能电感耦合等离子体发射光谱法(HP-ICP-OES)用于证明美国国家标准技术研究院(NIST)标准参考材料(SRM)1877氧化铍粉末中的Be质量分数。以95%置信水平表示的标准值和扩展的不确定度为(0.3576±0.0024)g / g。为了获得最佳结果,对于每个单独的HP-ICP-OES分析,将校准溶液的Be质量分数,Mn(内标)质量分数和基质组成与样品溶液进行仔细匹配。之所以使用这种“精确匹配”方法,是因为NIST的经验表明,它通常可以提高HP-ICP-OES分析的准确性和精密度。 NIST从未发表过这些评论。由于BeO的毒性和难以容纳非常细的粉末材料的原因,HP-ICP-OES分析的整套解决方案是由与NIST合作的实验室准备的,这些实验室拥有与材料安全合作所需的经验和设备。每个实验室都采用独特的消解方案。在准备好解决方案集之后,合作实验室将它们运送到NIST进行HP-ICP-OES分析。 NIST为合作实验室提供了解决方案准备工具包和电子表格,以帮助建立HP-ICP-OES结果与国际单位制(SI)的可追溯性,并实现精确匹配。由协作实验室制备的溶液组确定的四个单独的Be质量分数值之间的一致性为0.074%相对(平均值1s)。出色的协议可确保您对每种消化程序的耐用性以及经过认证的Be质量分数值充满信心。研究了使用完全匹配的特定证书的分析优势。结果表明,对于每次HP-ICP-OES分析,将标准品的基质成分与样品完全匹配对于获得各个Be质量分数值之间观察到的出色一致性至关重要,并且还有助于最大程度地减少标准值的偏差和不确定性。与以前的NIST研究不同,对于此特定认证,将标准品的Be和Mn质量分数与样品完全匹配似乎对数据影响很小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号