首页> 外文期刊>Analytical chemistry >Analysis of Melamine Cyanurate in Urine Using Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry
【24h】

Analysis of Melamine Cyanurate in Urine Using Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry

机译:基质辅助激光解吸/电离质谱分析尿液中的三聚氰胺三聚氰胺

获取原文
获取原文并翻译 | 示例
       

摘要

Matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) was applied to the direct analysis of melamine cyanurate (MC). The three commonly used MAL-DI matrixes, namely, alpha-cyano-4-hydroxycinnamic acid (CHCA), sinapinic acid (SA), and 2,5-dihydroxybenzoic acid (DHB), were able to desorb/ionize melamine from MC upon N_(2) laser irradiation, with CHCA showing the highest detection sensitivity in the positive mode. Only DHB and SA were able to desorb/ionize cyanuric acid from MC in the negative mode but with remarkably lower sensitivity. The method is able to detect melamine unambiguously from a small amount of MC (down to 12.5 (mu)g) spiked into urine and was successfully applied for the rapid and sensitive detection of melamine in urine stones/residues of the samples collected from patients clinically confirmed of having kidney stones associated with the consumption of melamine-tainted food products. The urine matrix resulted in interfering ion peaks and suppressed the ion intensity of melamine, while a cleanup process consisting of simply washing with water eliminated such interference and enhanced the ion intensity. The merit of the method is simplicity in sample preparation. The analytical time of the method for high-throughput analysis from the time of sample treatment to analysis is less than 7 minutes per sample, with sensitive detection of the presence of melamine in the urine stones/residues of the patient samples.
机译:基质辅助激光解吸/电离质谱(MALDI-MS)用于三聚氰胺氰尿酸酯(MC)的直接分析。三种常用的MAL-DI基质,即α-氰基-4-羟基肉桂酸(CHCA),芥子酸(SA)和2,5-二羟基苯甲酸(DHB)能够在MC上将三聚氰胺解吸/电离。 N_(2)激光照射,CHCA在正模式下显示出最高的检测灵敏度。只有DHB和SA能够以负模式从MC解吸/电离氰尿酸,但灵敏度显着降低。该方法能够从少量加标到尿液中的MC(低至12.5μg)中准确检测三聚氰胺,并成功地应用于临床上从患者收集的尿液结石/残留物中快速,灵敏地检测三聚氰胺。证实其肾脏结石与食用三聚氰胺污染的食品有关。尿液基质会干扰离子峰,并抑制三聚氰胺的离子强度,而仅用水冲洗就可以消除这种干扰并提高离子强度。该方法的优点是样品制备简单。从样品处理到分析的每个样品,高通量分析方法的分析时间少于7分钟,并且可以灵敏地检测到患者样品尿结石/残留物中三聚氰胺的存在。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号