首页> 外文期刊>Analytical chemistry >Rapid Determination of Plutonium Isotopes in Environmental Samples Using Sequential Injection Extraction Chromatography and Detection by Inductively Coupled Plasma Mass Spectrometry
【24h】

Rapid Determination of Plutonium Isotopes in Environmental Samples Using Sequential Injection Extraction Chromatography and Detection by Inductively Coupled Plasma Mass Spectrometry

机译:顺序注射萃取色谱法快速测定环境样品中的Is同位素并通过电感耦合等离子体质谱法进行检测

获取原文
获取原文并翻译 | 示例
       

摘要

This article presents an automated method for the rapid determination of ~(239)Pu and ~(240)Pu in various environmental samples. The analytical method involves the inline separation of Pu isotopes using extraction chromatography (TEVA) implemented in a sequential injection (SI) network followed by detection of isolated analytes with inductively coupled plasma mass spectrometry (ICP-MS). The method has been devised for the determination of Pu isotopes at environmentally relevant concentrations, whereby it has been successfully applied to the analyses of large volumes/amounts of samples, for example, 100-200 g of soil and sediment, 20 g of seaweed, and 200 L of seawater following analyte preconcentration. The investigation of the separation capability of the assembled SI system revealed that up to 200 g of soil or sediment can be treated using a column containing about 0.70 g of TEVA resin. The analytical results of Pu isotopes in the reference materials showed good agreement with the certified or reference values at the 0.05 significance level. Chemical yields of Pu ranged from 80 to 105percent, and the decontamination factors for uranium, thorium, mercury and lead were all above 10~(4). The duration of the in-line extraction chromatographic run was <1.5 h, and the proposed setup was able to handle up to 20 samples (14 mL each) in a fully automated mode using a single chromatographic column. The SI manifold is thus suitable for rapid and automated determination of Pu isotopes in environmental risk assessment and emergency preparedness scenarios.
机译:本文提出了一种自动测定各种环境样品中〜(239)Pu和〜(240)Pu的自动化方法。该分析方法包括使用萃取色谱法(TEVA)在线分离Pu同位素,该萃取色谱法是在顺序注射(SI)网络中实施的,然后通过电感耦合等离子体质谱(ICP-MS)检测分离的分析物。该方法已设计用于测定与环境相关浓度的Pu同位素,已成功应用于大量/大量样品的分析,例如100-200 g土壤和沉积物,20 g海藻,分析物预浓缩后再注入200升海水。对组装的SI系统的分离能力的研究表明,使用包含约0.70 g TEVA树脂的色谱柱可以处理多达200 g的土壤或沉积物。参考物质中Pu同位素的分析结果与0.05显着性水平的认证或参考值显示出良好的一致性。 Pu的化学收率在80%到105%之间,铀,or,汞和铅的去污因子都在10〜(4)以上。在线提取色谱运行的持续时间少于1.5小时,并且所提出的设置能够使用单个色谱柱以全自动模式处理多达20个样品(每个14 mL)。因此,SI歧管适用于在环境风险评估和应急准备情况下快速自动确定Pu同位素。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号