Scanning and Surface Alignment Considerations in Chemical Imaging with Desorption Electrospray Mass Spectrometry

机译:解吸电喷雾质谱法在化学成像中的扫描和表面对准注意事项

获取原文
获取原文并翻译 | 示例

摘要

The effects of surface scanning mode (raster vs unidirectional scanning) and the constancy of spray tip-to-surface and atmospheric sampling interface capillary-to-surface distances on chemical image quality using desorption electrospray ionization mass spectrometry were investigated. Unidirectional scanning was found to provide a spatially and a quantitatively more precise chemical image of the surface as compared to raster scanning. Maintaining constant spray tip-to-surface and atmospheric sampling interface capillary-to-surface distances during an imaging experiment was found to also be critical. An automation process was implemented using a custom image analysis software (HandsFree Surface Analysis) to keep these distances constant during the surface sampling experiment. Improved chemical image quality afforded through this software control was illustrated by imaging printed objects on normal copy paper.
机译:使用解吸电喷雾电离质谱法研究了表面扫描模式(光栅与单向扫描)以及喷雾头到表面和大气采样界面的毛细管到表面距离的恒定性对化学图像质量的影响。与光栅扫描相比,发现单向扫描可提供表面的空间和定量更精确的化学图像。发现在成像实验过程中保持恒定的喷嘴到表面和大气采样界面的毛细管到表面距离也很关键。使用自定义图像分析软件(HandsFree Surface Analysis)实施了自动化过程,以在表面采样实验期间将这些距离保持恒定。通过在普通复印纸上成像打印对象,可以说明通过此软件控件提供的改进的化学图像质量。