Capillary Column Switching Restricted-Access Media-Liquid Chromatography-Electrospray Ionization-Tandem Mass Spectrometry System for Simultaneous and Direct Analysis of Drugs in Biofluids

机译:毛细管柱切换受限访问介质-液相色谱-电喷雾电离-串联质谱系统可同时直接分析生物流体中的药物

获取原文
获取原文并翻译 | 示例
       

摘要

Capillary online restricted-access media-liquid chromatography-electrosprayionization-tandem mass spectrometry (RAM-LC-ESI-MS/MS) for direct analysis of drugs and metabolites spiked in biological fluids was developed. Using a column switching setup it was possible to perform effective sample preparation and analysis of raw biological fluids (plasma and urine) without matrix effects in the electrospray mass spectrometric detection step. The peak focusing efficiency of the extraction column was more effective in backflush compared to foreflush mode. The system was able to concentrate diminished samples of polar drugs and their metabolites reaching quantifiable results as low as 1 ng/mL utilizing a sample volume of only 333 nL of biofluids. New column hardware was developed to circumvent clogging problems experienced with plasma injections. The glass fiber filter frit, which is commonly used, was replaced with a short piece of 20 (mu)m i.d. fused silica capillary. The extraction columns were able to handle up to 60 injections and showed a high loading capacity, making the saturation of the MS detector the limiting factor on the linear dynamic range. The simultaneous separation and detection of 10 drugs and metabolites was obtained in 8 min of analysis, including the online sample preparation and enrichment step.
机译:开发了毛细管在线受限访问介质-液相色谱-电喷雾串联质谱(RAM-LC-ESI-MS / MS),用于直接分析加标在生物液体中的药物和代谢物。使用色谱柱切换装置,可以在电喷雾质谱检测步骤中进行有效的样品制备和原始生物流体(血浆和尿液)分析,而不会产生基质效应。与反吹模式相比,反吹中萃取柱的峰聚焦效率更高。该系统仅使用333 nL的生物流体样品就能浓缩极少的极性药物及其代谢物样品,从而可量化的结果低至1 ng / mL。开发了新的色谱柱硬件来避免血浆注入遇到的堵塞问题。常用的玻璃纤维滤芯用一小片20微米内径的玻璃片代替。熔融石英毛细管。萃取柱最多可处理60次进样,并显示出高载样量,这使得MS检测器的饱和度成为线性动态范围的限制因素。在8分钟的分析过程中,可以同时分离和检测10种药物和代谢物,包括在线样品制备和富集步骤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号