A fast, efficient, real-time method for the quantitative analysis of the peroxide explosive, TATP, is described. The method utilizes rapid ion-molecule reactions of chemical reagent ions with the vapor above solid TATP. The reactions of three reagent ions (H_(3)O~(+), O_(2)~(+), NO~(+)) were examined. Although all three ions exhibited a near-collision-rate reaction with TATP, only NO~(+) showed product ions that provide unequivocal evidence for a TATP-based explosive. The limit of detection of TATP in the gas phase is 10 ppb (4 X 10~(-10) mol L~(-1)).
机译:描述了一种快速,高效,实时的过氧化物炸药TATP定量分析方法。该方法利用化学试剂离子与高于固体TATP的蒸气的快速离子分子反应。检测了三种试剂离子(H_(3)O〜(+),O_(2)〜(+),NO〜(+))的反应。尽管所有三个离子均显示出与TATP接近的碰撞速率反应,但只有NO〜(+)显示出产物离子,这些离子为基于TATP的炸药提供了明确的证据。气相中TATP的检出限为10 ppb(4 X 10〜(-10)mol L〜(-1))。