Development of a Cryogen-Free Concentration System for Measurements of Volatile Organic Compounds

机译:开发用于测量挥发性有机化合物的无冰原浓度系统

获取原文
获取原文并翻译 | 示例
           

摘要

An innovative cryogen-free concentrator system for measurement of atmospheric trace gases at the parts per trillion level has been developed with detection by routinely used gas chromatographic methods. The first-generation system was capable of reaching a trapping temperature of -186 deg C, while the current version can reach -195 deg C. A Kleemenko cooler is used to create liquid nitrogen equivalent trapping conditions and eliminate the use of solid absorbents, a potential source of artifacts. The method utilizes dual-stage trapping with individual cold regions. The two stages are cooled to -20 and -175 deg C for water management and sample enrichment, respectively. Both stages house a Silonite-coated stainless steel sample loop; the second stage loop is filled with 1-mm-diameter glass beads, which provide an inert surface area for analyte concentration. In our application, the complete system employed four channels utilizing two flame ionization detectors, one electron capture detector, and a mass spectrometer. The system was automated for unattended operation and was deployed off the New England east coast on Appledore Island to measure a suite of ambient non-methane hydrocarbons, halocarbons, alkyl nitrates, and oxygenated volatile organic compounds during the International Consortium for Atmospheric Research on Transport and Transformation field campaign in summer 2004. This robust system quantified 98 ambient volatile organic compounds with precisions ranging from 0.3 to 15percent.
机译:开发了一种创新的无制冷剂浓缩器系统,用于测量万亿分之几的大气痕量气体,并通过常规使用的气相色谱法进行检测。第一代系统能够达到-186℃的捕集温度,而当前版本的系统可以达到-195℃。使用Kleemenko冷却器来创建液氮当量的捕集条件,并避免使用固体吸收剂,潜在的文物来源。该方法利用具有单个寒冷区域的双级捕集。将这两个阶段分别冷却至-20和-175摄氏度以进行水管理和样品富集。这两个平台都装有经硅藻土涂层的不锈钢样品环。第二级回路装有直径为1毫米的玻璃珠,可提供用于分析物浓缩的惰性表面积。在我们的应用中,整个系统采用了四个通道,其中使用了两个火焰离子化检测器,一个电子捕获检测器和一个质谱仪。该系统实现了无人值守的自动化操作,并在Appledore岛的新英格兰东海岸外部署,用于在国际运输和大气研究协会期间测量一组环境非甲烷碳氢化合物,卤代烃,烷基硝酸盐和氧化性挥发性有机化合物。 2004年夏季进行了转化现场活动。该功能强大的系统对98种环境挥发性有机化合物进行了定量,精度范围为0.3%至15%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号