Rectilinear Ion Trap Mass Spectrometer with Atmospheric Pressure Interface and Electrospray Ionization Source

机译:具有大气压接口和电喷雾电离源的直线型离子阱质谱仪

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A rectilinear ion trap (RIT) mass analyzer was incorporated into a mass spectrometer fitted with an electrospray ionization source and an atmospheric pressure interface. The RIT mass spectrometer, which was assembled in two different configurations, was used for the study of biological compounds, for which performance data are given. A variety of techniques, including the use of a balanced rf, elevated background gas pressure, automatic gain control, and resonance ejection waveforms with dynamically adjusted amplitude, were applied to enhance performance. The capabilities of the instrument were characterized using proteins, peptides, and pharmaceutical drugs. Unit resolution and an accuracy of better than m/z 0.2 was achieved for mass-to-charge (m/z) ratios up to 2000 Th at a scan rate of approx3000 amu/(charge(centre dot)s) while reduced scan rates gave greater resolution and peak widths of less than m/z 0.5 over the same range. The mass discrimination in trapping externally generated ions was characterized over the range m/z 190-2000 and an optimized low mass cutoff value of m/z 120-140 was found to give equal trapping efficiencies over the entire range. The radial detection efficiency was measured as a function of m/z ratio and found to rise from 35percent at low m/z values to more than 90percent for ions of m/z 1800. The way in which the ion trapping capacity depends on the dc trapping potential was investigated by measuring the mass shift due to space charge effects, and it was shown that low trapping potentials minimize space charge effects by increasing the useful volume of the device. The collision-induced dissociation (CID) capabilities of the RIT instrument were evaluated by measuring isolation efficiency as a function of mass resolution as well as measuring peptide CID efficiencies. Overall CID efficiencies of more than 60percent were easily reached, while isolation of an ion with unit resolution at m/z 524 was achieved with high rejection (>95percent) of the adjacent ions. The overall analytical capabilities of the ESI-RIT instrument were demonstrated with the analysis of a mixture of pharmaceutical compounds using multiple-stage mass spectrometry.
机译:将直线离子阱(RIT)质量分析仪安装到装有电喷雾电离源和大气压界面的质谱仪中。 RIT质谱仪以两种不同的配置组装而成,用于研究生物化合物,并给出了性能数据。为了提高性能,已应用了多种技术,包括使用平衡射频,升高的背景气压,自动增益控制以及具有动态调整幅度的共振喷射波形。使用蛋白质,肽和药物表征仪器的功能。当质荷比(m / z)高达2000 Th时,以大约3000 amu /(电荷(中心点)s)的扫描速率实现了单位分辨率和优于m / z 0.2的精度,同时降低了扫描速率在相同范围内得到的分离度更高,峰宽小于m / z 0.5。在m / z 190-2000范围内表征了捕获外部产生的离子的质量歧视,发现m / z 120-140的优化的低质量截留值在整个范围内具有相同的捕获效率。测得的径向检测效率是m / z比的函数,发现对于m / z 1800的离子,从低m / z值时的35%上升到90%以上。离子捕获能力的方式取决于dc通过测量由于空间电荷效应引起的质量转移来研究捕获势,结果表明低捕获势能通过增加器件的可用体积来最大程度地降低空间电荷效应。通过测量作为质量分辨率的函数的分离效率以及测量肽CID效率来评估RIT仪器的碰撞诱导解离(CID)能力。轻松实现超过60%的整体CID效率,同时以m / z 524的单位分辨率分离离子,并实现了对相邻离子的高排斥(> 95%)。 ESI-RIT仪器的整体分析能力通过使用多级质谱分析药物混合物来证明。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号