A novel aerosol chemiluminescent (CL) detector coupling to capillary electrophoresis (CE) for the detection of saccharides is reported. This CL detector is composed of a postcapillary nebulizer and porous alumina as catalyzer in quartz tube. The CL emission could be generated due to the catalyzing oxidization of saccharides on the surface of porous alumina. The saccharides such as sucrose, alpha-lactose, maltose, raffinose, galactose, xylose, and glucose with only weak UV absorbance can be successfully detected. The linear ranges of those saccharides are from 30-2000 to 50-2000 mg/L; relative standard deviations range from 2.1 to 3.7percent (200 mg/L, n velence 11). Compared with the traditional UV detector currently used in CE, this novel detector shows the advantage of high sensitivity to the compounds with only weak UV absorption. Thus, it could be an important supplement of CE detectors for UV-lacking compounds.
机译:新颖的气溶胶化学发光(CL)检测器耦合到毛细管电泳(CE)用于糖的检测。该CL检测器由毛细管后雾化器和多孔氧化铝作为石英管中的催化剂组成。由于多孔氧化铝表面上糖类的催化氧化,可能产生CL发射。可以成功检测到只有弱紫外线吸收的糖类,例如蔗糖,α-乳糖,麦芽糖,棉子糖,半乳糖,木糖和葡萄糖。这些糖的线性范围为30-2000至50-2000 mg / L;相对标准偏差范围为2.1%至3.7%(200 mg / L,nvelence 11)。与目前CE中使用的传统UV检测器相比,该新型检测器显示出对仅具有弱UV吸收的化合物具有高灵敏度的优势。因此,它可能是CE检测器中缺少紫外线的化合物的重要补充。