Revisiting stopping rules for iterative methods used in emission tomography.

机译:回顾放射断层成像中使用的迭代方法的停止规则。

获取原文
获取原文并翻译 | 示例
       

摘要

The expectation maximization algorithm is commonly used to reconstruct images obtained from positron emission tomography sinograms. For images with acceptable signal to noise ratios, iterations are terminated prior to convergence. A new quantitative and reproducible stopping rule is designed and validated on simulations using a Monte-Carlo generated transition matrix with a Poisson noise distribution on the sinogram data. Iterations are terminated at the solution which yields the most probable estimate of the emission densities while matching the sinogram data. It is more computationally efficient and more accurate than the standard stopping rule based on the Pearson's chi(2) test.
机译:期望最大化算法通常用于重建从正电子发射断层扫描正弦图获得的图像。对于具有可接受信噪比的图像,在收敛之前终止迭代。使用蒙特卡罗生成的过渡矩阵在正弦图数据上具有泊松噪声分布的仿真设计并验证了一种新的定量可再现的停止规则。迭代终止于解决方案,该解决方案在匹配正弦图数据的同时得出最可能的发射密度估计值。与基于Pearson's chi(2)检验的标准停止规则相比,它在计算效率和准确性方面更高。