首页> 外文期刊>Clinical anatomy: official journal of the American Association of Clinical Anatomists & the British Association of Clinical Anatomists >Effect of soft tissue thickness over the posterior border of the vertebral body and disc space on cage placement during posterior lumbar interbody fusion: A cadaveric study
【24h】

Effect of soft tissue thickness over the posterior border of the vertebral body and disc space on cage placement during posterior lumbar interbody fusion: A cadaveric study

机译:椎体后边界和椎间盘间隙上软组织厚度对后路腰椎椎体间融合术中笼放置的影响:尸体研究

获取原文
获取原文并翻译 | 示例
       

摘要

In the present study, we investigated whether there is a difference between visual depth (VD) and radiological image depth (RD) of cages (i.e., structural interbody support devices) placed in disc spaces during posterior lumbar interbody fusion and whether soft tissues covering the posterior border of the vertebral body and associated disc space are the cause of any observed differences. Using digital calipers, cages were inserted at a depth of 5 mm from the soft tissues covering the posterior border of the vertebral body and disc space under direct vision; this depth was defined as VD. After insertion, RD was measured in triplicate. The reliability of RD measurements was evaluated using an intraclass coefficient test. To identify the cause of differences between VD and RD, the thicknesses of soft tissues were measured microscopically. A total of 40 lumbar intervertebral disc spaces with cages were evaluated. The mean RD of cages was 3.12 mm, while the mean difference between the VD and RD of cages (DVRD) was 1.91 mm. On histological examination, the mean thickness of the soft tissue was 2.02 mm. Comparative analysis between histological values and DVRD showed no statistical difference (P = 1.14, 1.55, 0.06). There was a significant difference between VD and RD during cage placement, and soft tissue structure appeared to be responsible for the DVRD of inserted cages. Therefore, cages should be inserted deeper to account for differences between visual and radiological image depths. Clin. Anat. 25:1066-1073, 2012.
机译:在本研究中,我们调查了在腰椎后椎体间融合手术过程中放置​​在椎间盘间隙中的笼子(即结构体间支撑装置)的视觉深度(VD)和放射影像深度(RD)之间是否存在差异,以及是否覆盖了软组织椎体后缘和相关的椎间盘间隙是观察到差异的原因。使用数字卡尺,在直视下,将笼子插入覆盖椎体后缘和椎间盘间隙的距软组织5 mm的深度;该深度定义为VD。插入后,一式三份测量RD。使用组内系数测试评估RD测量的可靠性。为了确定VD和RD之间差异的原因,显微镜下测量了软组织的厚度。总共评估了40个带笼的腰椎间盘间隙。笼的平均RD为3.12毫米,而笼的VD和RD(DVRD)之间的平均差为1.91毫米。在组织学检查中,软组织的平均厚度为2.02mm。组织学值与DVRD的比较分析显示无统计学差异(P = 1.14,1.55,0.06)。在放置笼子的过程中,VD和RD之间存在显着差异,并且软组织结构似乎是插入笼子的DVRD的原因。因此,应将笼子更深地插入,以解决视觉和放射图像深度之间的差异。临床阿纳特25:1066-1073,2012。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号