首页> 外文期刊>Computers in Biology and Medicine >Comparison of FFT and adaptive ARMA methods in transcranial Doppler signals recorded from the cerebral vessels.
【24h】

Comparison of FFT and adaptive ARMA methods in transcranial Doppler signals recorded from the cerebral vessels.

机译:FFT和自适应ARMA方法在脑血管记录的经颅多普勒信号中的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

In this work, transcranial Doppler signals recorded from the temporal region of the brain on 35 patients were transferred to a personal computer by using a 16-bit sound card. Fast Fourier transform and adaptive auto regressive-moving average (A-ARMA) methods were applied to transcranial Doppler frequencies obtained from the middle cerebral artery in the temporal region. Spectral analyses were obtained to compare both methods for medical diagnoses. The sonograms obtained using A-ARMA method give better results for spectral resolution than the FFT method. The sonograms of A-ARMA method offer net envelope and better imaging, so that the determination of blood flow and brain pressure can be calculated more accurately. All diseases show higher resistance to flow than controls with no difference between males and females. Whereas values between disease classes differed, resistance within each class was remarkably constant.
机译:在这项工作中,通过使用16位声卡将35位患者的大脑颞区记录的经颅多普勒信号传输到个人计算机。快速傅里叶变换和自适应自回归移动平均(A-ARMA)方法应用于从颞中脑中动脉获得的经颅多普勒频率。获得光谱分析以比较两种医学诊断方法。与FFT方法相比,使用A-ARMA方法获得的超声图可提供更好的频谱分辨率结果。 A-ARMA方法的超声图可提供净包络和更好的成像,从而可以更准确地计算出血流量和脑压。所有疾病都显示出比对照组更高的抗流能力,男女之间没有差异。尽管疾病类别之间的值不同,但每个类别内的抗药性却相当恒定。