首页> 外文期刊>Clinica chimica acta: International journal of clinical chemistry and applied molecular biology >Human anti-animal antibodies interference in the Siemens Immulite chemiluminescent insulin immuno-assay: About one case.
【24h】

Human anti-animal antibodies interference in the Siemens Immulite chemiluminescent insulin immuno-assay: About one case.

机译:人抗动物抗体干扰Siemens Immulite化学发光胰岛素免疫测定:大约一例。

获取原文
获取原文并翻译 | 示例
           

摘要

We report the case of a 46 year old woman, working in a post-office, who has consulted her GP for unexplained asthenia and diffuse muscular pain. She felt stressed and tired but did not report any "real" feeling of sickness. In front of these non specific symptoms, the physician ordered a "general" blood analysis, including hormonal and serological tests. The results obtained by the Verviers Hospital laboratory, a general primary and secondary healthcare facility, were strictly normal, but an unexpected high level of fasting insulin at 147 muUI/mL (expected values: 6-27 muUI/mL) was found with the Immulite 2500 immunoanalyzer (Siemens Healthcare Diagnostics, Deerfield, IL).
机译:我们报告了一名46岁女性在邮局工作的案例,该女性曾就无法解释的乏力和弥漫性肌肉疼痛向其GP咨询。她感到压力和疲倦,但没有报告任何“真正的”疾病感。针对这些非特定症状,医生下令进行“常规”血液分析,包括激素和血清学检查。 Verviers医院实验室(一家普通的初级和二级医疗机构)获得的结果严格正常,但使用Immulite时发现空腹胰岛素的意外水平很高,为147 muUI / mL(预期值:6-27 muUI / mL)。 2500免疫分析仪(Siemens Healthcare Diagnostics,伊利诺伊州迪尔菲尔德)。