Systematic differences between BNP immunoassays: Comparison of methods using standard protocols and quality control materials

机译:BNP免疫测定之间的系统差异:使用标准方案和质量控制材料进行的方法比较

获取原文
获取原文并翻译 | 示例

摘要

Background: Recent studies suggested that there are marked systematic differences among BNP immunoassays. In this study we compared the BNP data and clinical results obtained with different immunoassays, including a new method (ST-AIA-PACK, TOSOH Corporation). Methods: BNP was measured on plasma-EDTA samples of healthy subjects (HS, n = 126) and patients with heart failure (HF, n = 31 NYHA I, II; n = 46 NYHA III, IV) using the ST-AIA-PACK and the Triage Biosite (Beckman Coulter) methods. Control samples distributed in the CardioOrmoCheck external quality assessment were also measured with TOSOH and the most used BNP immunoassays in Italy. Results: TOSOH method showed a good correlation (R = 0.976; n = 327) but a mean bias (-46.9%) compared to Triage Biosite. On the base of the results obtained in 10 samples of the CardioOrmoCheck study, TOSOH method showed a strict agreement with ADVIA Centaur, while it underestimated BNP in comparison with Triage (-52.5%) and ARCHITECT methods (-39.4%). The agreement of ST-AIA-PACK and Triage Biosite methods for classification of HF patients was tested using 100 ng/L of BNP; the positive agreement between methods was 65%, overall agreement was 73%. Conclusions: Our results confirm that there are marked differences in measured values among commercial methods for BNP assay.
机译:背景:最近的研究表明,BNP免疫测定之间存在明显的系统差异。在这项研究中,我们比较了BNP数据和使用不同免疫测定方法(包括一种新方法,ST-AIA-PACK,TOSOH Corporation)获得的临床结果。方法:使用ST-AIA-A对健康受试者(HS,n = 126)和心力衰竭(HF,n = 31 NYHA I,II; n = 46 NYHA III,IV)血浆-EDTA样品进行BNP测量。 PACK和Triage Biosite(贝克曼库尔特)方法。还使用TOSOH和意大利最常用的BNP免疫测定法测量了CardioOrmoCheck外部质量评估中分发的对照样品。结果:与Triage Biosite相比,TOSOH方法显示出良好的相关性(R = 0.976; n = 327),但平均偏倚(-46.9%)。根据CardioOrmoCheck研究的10个样本获得的结果,TOSOH方法与ADVIA Centaur表现出严格的一致性,而与Triage(-52.5%)和ARCHITECT方法(-39.4%)相比,它低估了BNP。使用100 ng / L的BNP测试了ST-AIA-PACK和Triage Biosite方法对HF患者分类的一致性。方法之间的阳性一致性为65%,总体一致性为73%。结论:我们的结果证实,商业化BNP测定方法的测量值存在显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号