BACKGROUND: Several automated instruments examining urine sediment have been introduced. We compared the performance of Sysmex UF-100 and Iris iQ200 with manual microscopy in urine sediment testing. METHODS: Four hundred and thirty-six urine samples were collected. The urine sediments were examined by manual microscopy and these 2 automated urinalysis systems. RESULTS: The within-run CVs for urine samples ranged from 3.4% to 22.3% for the iQ200, 1.6% to 24.2% for the UF-100 and 12.5% to 43.9% for manual microscopy. Between-run CVs on quality-control samples ranged from 6.1% to 32.4% for the iQ200 and 3.5% to 24.7% for the UF-100. The agreement between methods was good for red blood cells and white blood cells counts based on r values of 0.935 to 0.968. However, for epithelial cells, the values measured by different systems were poorly correlated (r=0.888-0.922). The Bland-Altman plot indicated a trend towards the automated cell count being greater than the manual microscopy as the epithelial cell countincreased. Casts were difficultly differentiated by 2 automated systems. CONCLUSIONS: These 2 automated urinalysis systems demonstrated good concordance with each other in urine sediment examination. The automated process could be used as a screening procedure but some manual microscopy was still necessary.
机译:背景:已经引进了几种检查尿沉渣的自动化仪器。我们比较了Sysmex UF-100和Iris iQ200与手动显微镜在尿沉渣测试中的性能。方法:收集了236份尿液样本。通过手动显微镜和这两个自动尿液分析系统检查尿液沉积物。结果:iQ200的尿液样品的运行CV范围从3.4%到22.3%,UF-100的运行内CV范围从1.6%到24.2%,手动显微镜的CV范围从12.5%到43.9%。对于iQ200,质控样品的批间CV范围为6.1%至32.4%,而对于UF-100,则为3.5%至24.7%。方法之间的一致性对于基于r值为0.935至0.968的红细胞和白细胞计数是有利的。然而,对于上皮细胞,通过不同系统测量的值之间的相关性很差(r = 0.888-0.922)。 Bland-Altman图表明随着上皮细胞计数的增加,自动细胞计数的趋势大于手动显微镜。演员表很难通过2个自动化系统进行区分。结论:这两个自动尿液分析系统在尿沉渣检查中表现出良好的一致性。自动化过程可以用作筛选程序,但仍需要一些手动显微镜检查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号