Development of a rapid, microplate-based kinetic assay for measuring adenosine deaminase activity in body fluids

机译:一种快速,基于微孔板的动力学测定法,用于测量体液中的腺苷脱氨酶活性

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Adenosine deaminase (ADA) catalyzes the deamination of adenosine to inosine. The activity of ADA in body fluids has clinical utility in the assessment of suspected tuberculosis. Methods: The conversion of adenosine to inosine was monitored in 96-well microplates as a continuous decrease at 265 nm for 15 min at ambient temperature. Analytical precision, sensitivity, linearity, accuracy, and enzyme stability were validated. Reference intervals were established from >120 tuberculosis-negative pleural, peritoneal, and cerebrospinal fluid samples. Results: The molar extinction coefficients of adenosine and inosine at 265 nm were 12,715 and 4,918 l/mol cm and their difference was used to calculate ADA activity. Maximum within-day imprecision was <11% and maximum total precision was <19%. Analytical sensitivity was 0.5 U/l and the assay was linear to 40 U/l. ADA recovery was 96-110% over an activity range of 11.7-25.3. U/l. ADA was stable for 1, 7 and 30 days at ~25 °C, 4-8 °C, and -20 °C storage, respectively. Upper reference limits were 9.4, 7.3, and 1.5. U/l for pleural, peritoneal, and cerebrospinal fluid, respectively. Conclusions: The microplate-based kinetic ADA assay has favorable performance characteristics. This method eliminates the need for assay calibration and allows 96 samples to be tested simultaneously.
机译:背景:腺苷脱氨酶(ADA)催化腺苷脱氨成肌苷。 ADA在体液中的活性在评估可疑结核病方面具有临床效用。方法:在96孔微孔板中监测腺苷向肌苷的转化,在环境温度下于265 nm处连续降低15分钟。分析精度,灵敏度,线性,准确性和酶稳定性得到验证。从> 120例阴性的胸膜,腹膜和脑脊液样本中确定参考间隔。结果:腺苷和肌苷在265 nm处的摩尔消光系数分别为12,715和4,918 l / mol cm,并将其差值用于计算ADA活性。日内最大不精确度为<11%,最大总精确度为<19%。分析灵敏度为0.5 U / l,测定线性为40 U / l。在11.7-25.3的活动范围内,ADA的回收率为96-110%。单位ADA分别在〜25°C,4-8°C和-20°C的存储下稳定1、7和30天。参考上限为9.4、7.3和1.5。胸膜,腹膜和脑脊液分别为U / l。结论:基于微孔板的动力学ADA测定具有良好的性能特征。这种方法消除了对测定标定的需要,并允许同时测试96个样品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号