The first calibration-free Near-Patient-Testing instrument (NPT7) for blood gases, pH and oximetry has been developed. With cartridges of 30 single-use cuvettes, the NPT7 needs no preparation prior to sample aspiration, no manual calibration, and no maintenance apart from paper and cartridge changes and regulatory quality control. Each cuvette measures pCO2, pO2, pH, total hemoglobin (ctHb), oxygen saturation (sO2), fractions of carboxyhemoglobin (FCOHb) and methemoglobin (FMetHb) on 95 microl whole blood with a 110-s measuring cycle. The measurement principles are as follows: pCO2-three-wavelength infrared spectroscopy of dissolved CO2; pO2-measurement of O2-induced changes in the decay time of phosphorescence; pH-the absorbance spectra change of an azo-dye color indicator; and oximetry is performed with a 128-wavelength spectrophotometer. We determined the within and between instrument variations with tonometered whole blood on seven prototype instruments, using between one and five control levels per analyte. The 95% analytical performance limits: +/-(/Bias/ +2 xS(T)) in the NPT7 instrument matched the analytical performance criteria for the measured quantities as defined by AACC guidelines. The application of these optical measuring methods for blood gases, pH and oximetry in single-use devices introduces a new concept into point-of-care testing (POCT), where preanalytical activities otherwise associated with instrument preparation are eliminated.
机译:已开发出第一台用于血气,pH和血氧饱和度的免校准近患者测试仪(NPT7)。 NPT7带有30个一次性使用的试管,无需进行任何准备工作即可进行样品抽吸,手动校准以及维护,除了更换纸张和墨盒以及进行法规质量控制外。每个比色皿以110秒的测量周期测量95微升全血中的pCO2,pO2,pH,总血红蛋白(ctHb),氧饱和度(sO2),羧基血红蛋白(FCOHb)和高铁血红蛋白(FMetHb)的分数。测量原理如下:溶解的二氧化碳的pCO2-三波长红外光谱法; pO2测量O2引起的磷光衰减时间的变化; pH-偶氮染料颜色指示剂的吸收光谱变化;用128波长分光光度计进行血氧测定。我们使用七种原型仪器对眼压计全血确定了仪器内部和仪器之间的差异,每种分析物使用一到五个对照水平。 95%的分析性能极限:NPT7仪器中的+/-(/ Bias / +2 xS(T))符合AACC指南定义的测量量的分析性能标准。这些光学测量方法在一次性设备中用于血气,pH和血氧测定法的应用为护理点测试(POCT)引入了新概念,从而消除了与仪器准备相关的分析前活动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号