首页> 外文期刊>Clinical and laboratory haematology >Comparison of the reticulocyte mode of the Abx Pentra 120 Retic, Coulter General-S, Sysmex SE 9500, Abbott CD 4000 and Bayer Advia 120 haematology analysers in a simultaneous evaluation.
【24h】

Comparison of the reticulocyte mode of the Abx Pentra 120 Retic, Coulter General-S, Sysmex SE 9500, Abbott CD 4000 and Bayer Advia 120 haematology analysers in a simultaneous evaluation.

机译:同时评估Abx Pentra 120 Retic,Coulter General-S,Sysmex SE 9500,Abbott CD 4000和Bayer Advia 120血液分析仪的网状细胞模式的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

The Abx Pentra 120 Retic, Coulter General-S, Sysmex SE 9500, Abbott Cell Dyn 4000 and Bayer Advia 120 were evaluated simultaneously in a general hospital laboratory. Linearity, precision, sample stability, carry-over and comparability of the reticulocyte mode were determined following International Council for Standardization in Haematology guidelines for the evaluation of blood cell analysers. All analysers showed good results for dilution, stability and carry-over testing. The between-batch coefficient of variation of the General-S was high compared to the other analysers evaluated. Multiple correlation studies showed good agreement for all analysers in the normal and high reticulocyte range, with correlation coefficients above 0.7. Multiple correlation studies for reticulocytopenic samples (< 15.109/l) were less satisfactory, with a wider range of correlation coefficients (r-values 0.0-0.9). Overall, the General-S, SE 9500 and Advia 120 gave lower reticulocyte counts than the Pentra 120 Retic and CD 4000. Reagent costs were also evaluated. Reagent consumption was close to the manufacturers' specifications for the SE 9500 (Search reagent), CD 4000 (CD Retic) and Advia 120 (Retics) but was higher than stated for the Pentra 120 Retic (Retix), General-S (Retic kit) and SE 9500 (Sheath reagent). Our results show that these new generation haematology analysers will meet the needs of hospital laboratories for reliable and cost-effective reticulocyte counting.
机译:在综合医院实验室同时评估了Abx Pentra 120 Retic,Coulter General-S,Sysmex SE 9500,Abbott Cell Dyn 4000和Bayer Advia 120。网织红细胞模式的线性,精密度,样品稳定性,残留和可比性是根据国际血液学标准化委员会血液细胞分析仪评估指南确定的。所有分析仪在稀释,稳定性和残留测试方面均显示出良好的结果。与其他评估分析仪相比,General-S的批次间变异系数很高。多项相关研究表明,所有分析仪在正常和高网织红细胞范围内均具有良好的一致性,相关系数高于0.7。网状细胞减少症样品的多重相关性研究(<15.109 / l)不令人满意,相关系数范围更广(r值0.0-0.9)。总的来说,与Pentra 120 Retic和CD 4000相比,General-S,SE 9500和Advia 120的网织红细胞计数更低。还评估了试剂成本。试剂消耗量接近SE 9500(搜索试剂),CD 4000(CD Retic)和Advia 120(Retic)的制造商规格,但高于Pentra 120 Retic(Retix),General-S(Retic试剂盒)的规定。 )和SE 9500(护套试剂)。我们的结果表明,这些新一代血液学分析仪将满足医院实验室对网织红细胞计数可靠且具有成本效益的需求。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号