Electrical conductivity of a solution is a property that can be easily determined through the measurement of a conductivity probe. The present study demonstrates the measurements of electrical conductivity for two ternary solutions: glycerol/sodium chloride/water and ethylene glycol/sodium chloride/water. When the concentration of sodium chloride to water ratio (R) is fixed, the existence of either glycerol or ethylene glycol, both cryoprotec-live agents (CPAs), can be quantitatively determined by their depressive influence on electrical conductivity of the solution. The measurements were performed on solutions with a set of 10 different concentrations of CPAs, ranging from 3.2% to 50% (v/v), along with five ratios of NaCl/water solutions. Equations to fit the experimental measurements were devised to characterize the relations among electrical conductivity, CPAs concentration, and R. A conductivity meter used in this study required <5 s to read the solution's electrical conductivity, which is faster than the measurement using osmometry method. The charts of ternary solutions associated with their electrical conductivity and concentrations make it especially useful for monitoring the cryopreservation pro-cesses, including addition and removal of CPAs, to prevent osmotic damages to biological samples.
机译:溶液的电导率是可以通过电导率探针的测量容易确定的特性。本研究证明了两种三元溶液的电导率测量:甘油/氯化钠/水和乙二醇/氯化钠/水。当氯化钠与水的比例(R)的浓度固定时,甘油和乙二醇(两种低温保护活性剂(CPA))的存在可以通过其对溶液电导率的抑制作用来定量确定。测量是在具有10种不同浓度的CPA(范围从3.2%至50%(v / v))以及五种NaCl /水溶液比例的溶液上进行的。设计了适合实验测量的方程式,以表征电导率,CPA浓度和R之间的关系。本研究中使用的电导率仪需要不到5 s的时间读取溶液的电导率,这比使用渗透压法的测量更快。与它们的电导率和浓度相关的三元溶液图使其特别适用于监测低温保存过程,包括添加和去除CPA,以防止对生物样品的渗透破坏。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号