Compton spectroscopy for rotation-mode computed tomography.

机译:康普顿光谱用于旋转模式计算机断层扫描。

获取原文
获取原文并翻译 | 示例
       

摘要

In computed tomography (CT) systems, it is desirable to know the X-ray energy spectra for various applications, including medical CT imaging, and diagnostic field and heavy ion therapy. However, because of the restricted space, the only practical solution is to use Compton spectroscopy, where the incident spectrum is inferred from the scattered spectrum. The geometry of the scatterer and its position within the CT can affect the spectrum of the secondary beam, making it difficult to determine the primary spectrum during operation of the CT system. A modified Compton spectrometer is described that allows measurement of the X-ray energy spectra during operation, and most importantly, in rotation mode. The geometry of the scatterer was optimized to reduce the energy broadening of the secondary beam. The performance of the system was evaluated by comparing the reconstructed exposure to that measured directly using an ion chamber. [ABSTRACT FROM AUTHOR]
机译:在计算机断层摄影(CT)系统中,期望了解各种应用的X射线能谱,包括医学CT成像,诊断领域和重离子治疗。但是,由于空间有限,唯一可行的解​​决方案是使用康普顿光谱法,其中从散射光谱推断入射光谱。散射体的几何形状及其在CT中的位置会影响副光束的光谱,从而在CT系统运行期间难以确定主光谱。描述了一种改进的康普顿光谱仪,其允许在操作期间并且最重要的是在旋转模式下测量X射线能谱。优化了散射体的几何形状,以减少副光束的能量展宽。通过将重建的暴露量与使用离子室直接测量的暴露量进行比较,可以评估系统的性能。 [作者的摘要]