Quantitative investigation of the effects of the scanning parameters in the digitization of EBT and EBT2 Gafchromic film dosimetry with flatbed scanners

机译:平板扫描仪对EBT和EBT2全色薄膜剂量学数字化过程中扫描参数的影响进行定量研究

获取原文
获取原文并翻译 | 示例
       

摘要

With increasing popularity and complexity of intensity-modulated radiation therapy (IMRT) delivery modalities including regular and arc therapies, there is a growing challenge for validating the accuracy of dose distributions. Gafchromic films have superior characteristics for dose verification over other conventional dosimeters. In order to optimize the use of Gafchromic films in clinical IMRT quality assurance procedures, the scanning parameters of EBT and EBT2 films with a flatbed scanner were investigated. The effects of several parameters including scanning position, orientation, uniformity, film sensitivity and optical density (OD) growth after irradiation were quantified. The profiles of the EBT and EBT2 films had a noise level of 0.6% and 0.7%, respectively. Considerable orientation dependence was observed and the scanner value difference between landscape and portrait modes were about 12% and 10% for EBT and EBT2 films, respectively. The highest response sensitivity was observed using digitized red color images of the EBT2 film scanned with landscape mode. The total system non-uniformity composed of contributions from the film and the scanner was less than 1.7%. OD variations showed that EBT gray scale grew slower, however, reached higher growth values of 15% when compared with EBT2 gray scale which grew 12% after a long time (480 hours) post-irradiation. The EBT film using the red color channel showed the minimal growth where OD increased up to 3% within 3 days after irradiation, and took one week to stabilize.
机译:随着包括常规疗法和电弧疗法在内的强度调制放射疗法(IMRT)输送方式的日益普及和复杂性,验证剂量分布的准确性面临着越来越大的挑战。与其他常规剂量计相比,全色薄膜具有出色的剂量验证特性。为了在临床IMRT质量保证程序中优化Gafchromic胶片的使用,研究了使用平板扫描仪扫描EBT和EBT2胶片的扫描参数。量化了几个参数的影响,包括扫描位置,方向,均匀性,胶片灵敏度和辐照后光密度(OD)的增长。 EBT和EBT2薄膜的轮廓噪声水平分别为0.6%和0.7%。观察到相当大的取向依赖性,并且对于EBT和EBT2胶片,横向和纵向模式之间的扫描仪值差异分别约为12%和10%。使用横向模式扫描的EBT2胶片的数字化红色图像可观察到最高的响应灵敏度。由胶片和扫描仪产生的总系统不均匀度小于1.7%。 OD的变化表明EBT灰度的增长较慢,但是与EBT2灰度相比,后者的辐照时间长(480小时)后增长了12%,达到了15%的更高增长值。使用红色通道的EBT膜显示出最小的增长,在辐照后3天内OD增至3%,并且需要一周时间才能稳定下来。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号