The Effect of Dose Grid Resolution on Dose Volume Histograms for Slender Organs at Risk during Pelvic Intensity-modulated Radiotherapy

机译:骨盆强度调制放射治疗中剂量网格分辨率对处于危险中的细长器官剂量体积直方图的影响

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: There are enduring uncertainties regarding the optimal dose grid resolution for use with pelvic intensity-modulated radiotherapy (IMRT) plans in which the adjacent organs at risk are slender and transect the field edge. Therefore, this study evaluated the effect of dose grid resolution on bladder wail dose-volume histogram (DVH) calculations for prostate IMRT plans.Materials and Methods: The planning computed tomography scans and clinical plans from 15 prostate cancer patients were included in this analysis. For each study computed tomography, the entire inner and outer bladder surfaces were delineated. Nine versions of the clinical plan were created, varying interval between the dose grid calculation points uniformly in three dimensions, whereas all other plan parameters were kept constant. The dose grid increments tested were 1-10 mm. The plans were recalculated and the bladder wall DVH compared against the study benchmark (1 mm grid).
机译:目的:对于盆腔强度调制放射治疗(IMRT)计划使用的最佳剂量网格分辨率存在持久的不确定性,在该计划中,处于危险中的相邻器官细长并横切视野边缘。因此,本研究评估了剂量网格分辨率对前列腺IMRT计划膀胱壁剂量体积直方图(DVH)计算的影响。材料与方法:这项分析包括15例前列腺癌患者的计划计算机断层扫描和临床计划。对于每项研究的计算机断层扫描,都描绘了整个膀胱内外表面。创建了九个版本的临床计划,剂量网格计算点之间的间隔在三个维度上均一变化,而所有其他计划参数均保持不变。测试的剂量网格增量为1-10 mm。重新计算了计划,并将膀胱壁DVH与研究基准(1毫米网格)进行了比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号