A new electrode design to improve outcomes in the treatment of chronic non-healing wounds in diabetes.

机译:一种新的电极设计,可改善糖尿病慢性不愈合伤口的治疗效果。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: Chronic wounds are life-threatening in people with diabetes. Some studies show that electrical stimulation (ES) can help wounds heal, while others do not. But, ES is usually applied using a two-electrode system, where current distribution is greatest in the center line between the electrodes. In the present study, a three-electrode system (three-channel ES) was developed. Current dispersion on the skin and in the quadriceps muscle was compared between the conventional two-electrode and three-electrode systems in controls and tested for its ability to heal chronic wounds in people with diabetes. METHODS: In controls, current was delivered via a biphasic sine wave at a frequency of 30 Hz and pulse width of 100 microseconds. Stimulation electrodes 5 cm x 5 cm and 5 cm x 10 cm were placed at 10 cm and 15 cm separation distances above the quadriceps muscle. Skin currents were measured using five pairs of surface electrodes positioned in five separate locations on the skin. Muscle currents were measured using three pairs of needle electrodes positioned in three different locations in the muscle belly. In chronic wounds in eight subjects with diabetes, stimulation was applied for 1 month, and healing and blood flow were measured. RESULTS: Current during three-channel ES was dispersed more evenly and more deeply than with conventional two-channel ES (P < 0.05). In wounds, there was almost complete healing in 1 month, and current was uniform in the wound. CONCLUSIONS: Three-channel ES is more effective than two-channel ES in terms of better current dispersion across the skin and penetration into tissue and will probably be better for wound healing.
机译:背景:慢性伤口威胁着糖尿病患者的生命。一些研究表明,电刺激(ES)可以帮助伤口愈合,而其他研究则不能。但是,ES通常使用两电极系统应用,其中电流分布在电极之间的中心线最大。在本研究中,开发了一种三电极系统(三通道ES)。比较了对照组中传统的两电极和三电极系统在皮肤和股四头肌上的电流散布,并测试了其治愈糖尿病患者慢性伤口的能力。方法:在对照中,电流通过双相正弦波以30 Hz的频率和100微秒的脉冲宽度传递。将刺激电极5 cm x 5 cm和5 cm x 10 cm放置在股四头肌上方10 cm和15 cm的分隔距离处。使用位于皮肤上五个不同位置的五对表面电极测量皮肤电流。使用位于肌肉腹部三个不同位置的三对针形电极测量肌肉电流。在八名糖尿病患者的慢性伤口中,施加刺激1个月,并测量其愈合情况和血流量。结果:与常规的两通道ES相比,三通道ES期间的电流分布更均匀,更深(P <0.05)。在伤口中,伤口在1个月内几乎完全愈合,伤口电流均匀。结论:就更好的电流在皮肤上的分散和渗透到组织中而言,三通道ES比两通道ES更有效,并且可能对伤口愈合更好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号