首页> 外文期刊>Journal of vascular and interventional radiology: JVIR >Indirect computed tomography venography of the lower extremities using single-source dual-energy computed tomography: Advantage of low-kiloelectron volt monochromatic images
【24h】

Indirect computed tomography venography of the lower extremities using single-source dual-energy computed tomography: Advantage of low-kiloelectron volt monochromatic images

机译:使用单源双能计算机断层摄影术对下肢进行间接计算机断层摄影术静脉造影:低千伏电子伏特单色图像的优势

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To study the performance of dual-energy indirect computed tomography (CT) venography from single-source dual-energy CT in the assessment of lower extremity deep venous thrombosis (DVT). Materials and Methods: In a retrospective study, 110 patients suspected to have pulmonary embolism (PE) underwent dual-energy CT venography on a single-source dual-energy CT scanner as a part of CT pulmonary angiography protocol at 3 minutes after injection of contrast material. Two radiologists evaluated 50-kiloelectron volt (keV) and 70-keV monochromatic images reconstructed from a dual-energy CT scan for image quality, image noise, venous contrast, and confidence level in interpretation for DVT using a scale of 1-5. In addition, a combined 50-keV and 70-keV data set was assessed for confidence level in image interpretation. Attenuation, contrast-to-noise ratio (CNR), and objective noise were measured in bilateral common femoral and popliteal veins. Data were analyzed using Student t test and Wilcoxon rank sum test. Radiation dose was measured for dual-energy CT venography protocol. Results: A diagnosis of DVT was made in 8 of 110 patients (7.27%). The subjective image quality was comparable between 50-keV and 70-keV images (4.3 vs 4.5; P > .05). The subjective venous contrast opacification (4.7 vs 3.5; P = .0036) and confidence (4.8 vs 3.9; P = .0028) in image interpretation were superior at 50 keV. Confidence level for interpretation on combined 50-keV and 70-keV series (score 4.7) was similar to that for 50-keV series (score 4.8). Compared with 70-keV data, 50-keV data yielded 90% increase in intravascular CT attenuation (207.4 Hounsfield units [HU] ± 39.0 vs 106.8 HU ±7.6; P <.0001) and higher CNR (10.7 ± 4.07 vs 7.2 ± 4.1; P = .0001) of the deep veins. However, objective noise at 50 keV was higher (14.8 HU vs 6.5 HU; P = .0031). Because of inadequate contrast opacification, 6% of CT venography studies were deemed suboptimal for rendering a diagnostic interpretation on 70-keV images, but these images were considered acceptable at 50 keV. The mean effective radiation dose for the dual-energy CT venography examination was 4.2 mSv. Conclusions: Optimal image quality with substantially higher venous attenuation is provided by 50-keV monochromatic images from dual-energy CT venography acquisition compared with 70-keV images. The 50-keV monochromatic images increase the confidence in the image interpretation of DVT and decrease the number of indeterminate studies.
机译:目的:研究单源双能CT的双能间接计算机断层扫描(CT)静脉造影在评估下肢深静脉血栓形成(DVT)中的性能。材料和方法:在一项回顾性研究中,在注入造影剂后3分钟,对110名怀疑患有肺栓塞(PE)的患者在单源双能CT扫描仪上进行了双能CT静脉造影,作为CT肺血管造影方案的一部分。材料。两位放射科医生评估了通过双能CT扫描重建的50千电子伏特(keV)和70 keV单色图像的图像质量,图像噪声,静脉对比度和DVT解释的置信度,使用1-5级。此外,评估了组合的50keV和70keV数据集的图像解释置信度。测量双侧股总和pop静脉的衰减,对比噪声比(CNR)和客观噪声。使用Student t检验和Wilcoxon秩和检验对数据进行分析。测量双能CT静脉造影方案的辐射剂量。结果:110例患者中有8例诊断为DVT(7.27%)。 50 keV和70 keV图像之间的主观图像质量相当(4.3 vs 4.5; P> .05)。在50 keV时,图像解释的主观静脉对比度浑浊(4.7 vs 3.5; P = .0036)和置信度(4.8 vs 3.9; P = 0.0028)更好。 50 keV和70 keV组合的解释的置信度(得分4.7)与50 keV系列的解释的可信度(得分4.8)相似。与70-keV数据相比,50-keV数据使血管内CT衰减增加了90%(207.4 Hounsfield单位[HU]±39.0 vs 106.8 HU±7.6; P <.0001)和更高的CNR(10.7±4.07 vs 7.2±4.1) ; P = .0001)的深静脉。但是,在50 keV时的客观噪声较高(14.8 HU对6.5 HU; P = 0.0031)。由于对比度浑浊不充分,对于在70 keV图像上进行诊断解释,CT静脉造影研究被认为次优,但这些图像在50 keV时被认为可以接受。双能CT静脉造影检查的平均有效辐射剂量为4.2 mSv。结论:与70keV图像相比,双能CT静脉成像采集的50keV单色图像可提供最佳的静脉衰减图像质量。 50 keV单色图像增加了DVT图像解释的可信度,并减少了不确定研究的次数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号