Penetration of Gunther Tulip filter struts through an introducer sheath: case report and safety concerns.

机译:Gunther Tulip过滤器支杆穿过导引器护套的穿透:病例报告和安全问题。

获取原文
获取原文并翻译 | 示例

摘要

The authors present a patient who had a Gunther-Tulip inferior vena cava filter placed under fluoroscopic guidance. The filter struts were seen to penetrate through the introducer sheath at deployment. This was believed to be secondary to a kink in the sheath at the site of venous entry and was due to the patient's nuchal obesity as well as his inability to turn his head to the opposite side. The introducer sheath was then placed through a long reinforced metal sheath through which the filter was then placed without complication.
机译:作者介绍了一名在荧光镜引导下放置了Gunther-Tulip下腔静脉滤器的患者。看到过滤器撑杆在展开时穿透导引器护套。据认为,这是由于静脉入口部位的鞘内扭结引起的,是由于患者的颈部肥胖症以及他无法将头转向另一侧。然后将导引器护套穿过长的增强金属护套,然后将过滤器毫无问题地通过该护套放置。