首页> 外文期刊>Journal of Veterinary Diagnostic Investigation >Comparison of two methods for blood lead analysis in cattle: graphite-furnace atomic absorption spectrometry and LeadCareReg. II system.
【24h】

Comparison of two methods for blood lead analysis in cattle: graphite-furnace atomic absorption spectrometry and LeadCareReg. II system.

机译:两种牛血铅分析方法的比较:石墨炉原子吸收光谱法和LeadCareReg。 II系统。

获取原文
获取原文并翻译 | 示例
           

摘要

The current study compared the LeadCareReg. II test kit system with graphite-furnace atomic absorption spectrometry for blood lead (Pb) analysis in 56 cattle accidentally exposed to Pb in the field. Blood Pb concentrations were determined by LeadCare II within 4 hr of collection and after 72 hr of refrigeration. Blood Pb concentrations were determined by atomic absorption spectrometry, and samples that were coagulated (n=12) were homogenized before analysis. There was strong rank correlation (R2=0.96) between atomic absorption and LeadCare II (within 4 hr of collection), and a conversion formula was determined for values within the observed range (3-91 mcg/dl, although few had values >40 mcg/dl). Median and mean blood pb concentrations for atomic absorption were 7.7 and 15.9 mcg/dl, respectively; for LeadCare II, medians were 5.2 mcg/dl at 4 hr and 4.9 mcg/dl at 72 hr, and means were 12.4 and 11.7, respectively. LeadCare II results at 4 hr strongly correlated with 72 hr results (R2=0.96), but results at 72 hr were lower (P<0.01). There was no significant difference between coagulated and uncoagulated samples run by atomic absorption. Although there have been several articles that compared LeadCare with other analytical techniques, all were for the original system, not LeadCare II. The present study indicated that LeadCare II results correlated well with atomic absorption over a wide range of blood Pb concentrations and that refrigerating samples for up to 72 hr before LeadCare II analysis was acceptable for clinical purposes.
机译:当前的研究比较了LeadCareReg。 II带有石墨炉原子吸收光谱法的测试套件系统,用于在野外意外接触56头牛的56头牛中的血铅(Pb)分析。在收集的4小时内和冷藏72小时后,通过LeadCare II测定血铅浓度。通过原子吸收光谱法测定血中铅的浓度,并在分析前将凝结的样品(n = 12)均质化。原子吸收与LeadCare II(收集后4小时之内)之间存在强的等级相关性( R 2 = 0.96),并确定了所观察值内的转换公式范围(3-91 mcg / dl,尽管几乎没有值> 40 mcg / dl)。原子吸收的中值和平均血铅浓度分别为7.7和15.9 mcg / dl。 LeadCare II的中位数在4小时时为5.2 mcg / dl,在72小时时为4.9 mcg / dl,平均值分别为12.4和11.7。 LeadCare II在4小时的结果与72小时的结果高度相关( R 2 = 0.96),但在72小时的结果则较低( P < 0.01)。通过原子吸收进行的凝结和未凝结样品之间没有显着差异。尽管有几篇文章将LeadCare与其他分析技术进行了比较,但所有文章都是针对原始系统的,而不是针对LeadCare II。本研究表明LeadCare II的结果与在广泛的血液Pb浓度范围内的原子吸收密切相关,并且在LeadCare II分析用于临床目的之前可以将样品冷冻长达72小时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号