Modern hip implants typically feature modular heads, which allow for easy exchange and removal from the femoral stem at the time of revision. However, owing to fretting, corrosion, or cold welding, the modular head may be difficult or impossible to separate from the underlying trunnion, especially if the implant has titanium interfaces between the head and the stem. We have repeatedly encountered difficulty removing the titanium sleeve adapter in the M(2)a-Magnum(TM) implant. Although the manufacturer warns about this complication and cases with these difficulties have been reported to the United States FDA, we believed this topic is important to study, because the frequency of difficulties in head removal is unknown and the complications related to this event have not been characterized.
机译:现代的髋关节植入物通常具有模块化的头部,可在翻修时方便地从股骨柄更换和取出。但是,由于微动,腐蚀或冷焊,模块化头可能很难或不可能与下面的耳轴分离,尤其是当植入物在头和茎之间具有钛界面时。我们屡屡遇到在M(2)a-MagnumTM植入物中移除钛套适配器的困难。尽管制造商警告了这种并发症,并且已经向美国FDA报告了这些困难的病例,但是我们认为该主题很重要,因为拔除头顶的困难的频率未知,并且尚未发现与此事件相关的并发症表征。