The quick-scanning XAFS (QXAFS) method is achieved at the BL14W1 XAFS beamline at the Shanghai Synchrotron Radiation Facility based on the EPICS and LabVIEW systems. This is realised by the unprecedented use of LabVIEWs data logging and supervisory control module for communication with EPICS in synchrotron radiation facilities. A fine QXAFS spectrum with an energy range of 1.2 keV at the Cu K-edge has been collected in 2 s with stable beam position and the data quality is comparable with that of the step-mode XAFS spectrum. Analog-to-digital converter and double-crystal monochromator set-ups have been optimized in order to acquire optimal parameters for the QXAFS experiments. Signal-to-noise ratios of these spectra have been calculated in order to estimate the importance of these parameters.
机译:快速扫描XAFS(QXAFS)方法是基于EPICS和LabVIEW系统在上海同步加速器辐射设施的BL14W1 XAFS光束线上实现的。这是通过在同步加速器辐射设施中前所未有地使用LabVIEW数据记录和监督控制模块与EPICS进行通信来实现的。在2 s内以稳定的光束位置收集了在Cu K边缘处能量范围为1.2 keV的精细QXAFS光谱,其数据质量与步进模式XAFS光谱相当。为了获得QXAFS实验的最佳参数,模数转换器和双晶体单色仪的设置已得到优化。为了估计这些参数的重要性,已经计算了这些光谱的信噪比。