X-ray imaging detectors with an identical phosphor and a CCD chip but employing lens- and fiber-coupling between them have been compared. These are designed for X-ray imaging experiments, especially computed tomography, at the medium-length beamline at the SPring-8 synchrotron radiation facility. It was found that the transmittance of light to the CCD is about four times higher in the fiber-coupled detector. The uniformity of response in the lens-coupled detector has a global shading of up to 40%, while pixel-to-pixel variation owing to a chicken-wire pattern was dominant in the fiber-coupled detector. Apart from the higher transmittance, the fiber-coupled detector has a few characteristics that require attention when it is used for computed tomography, which are browning of the fiber, discontinuity in the image, image distortion, and dark spots in the chicken-wire pattern. Thus, it is most suitable for high-speed tomography of samples that tend to deform, for example biological and soft materials.
机译:比较了具有相同荧光粉和CCD芯片但在它们之间采用透镜和光纤耦合的X射线成像检测器。这些设计用于在SPring-8同步加速器辐射设施的中等长度光束线上进行X射线成像实验,尤其是计算机断层扫描。发现在光纤耦合检测器中,光到CCD的透射率大约高四倍。透镜耦合检测器的响应均匀性具有高达40%的整体阴影,而在光纤耦合检测器中,由于鸡丝图案引起的像素间差异是主要的。光纤耦合检测器除了具有较高的透射率外,还具有一些在进行计算机断层扫描时需要注意的特性,例如光纤褐变,图像不连续,图像畸变以及网状图案中的黑点。 。因此,它最适合于容易变形的样品(例如生物材料和软材料)的高速层析成像。