Estimated Blood Pressure Algorithm for a Wrist-wearable Pulsimeter Using Hall Device

机译:使用霍尔装置的腕式脉搏计的血压估算算法

获取原文
获取原文并翻译 | 示例
           

摘要

In order to measure the blood pressure (BP) by using the pulse wave obtained from the radial artery, we developed a cuffless wrist-wearable pulsimeter with a Hall device. Based on an analysis of the non-pressurized pulse wave obtained from the Hall device, which detects minute changes in the magnetic field of a permanent magnet, a new algorithm for estimating the BP in oriental diagnostic medicine is proposed. Using the pulse wave data obtained from a clinical trial involving 13 subjects, we derived an equation for the regression analysis. This equation can be used to determine the correlation factor. The standard deviation of the estimated and measured values for high blood pressure was 3.5 mmHg and that for low blood pressure was 3.8 mmHg, which were close to the BP International Standard.
机译:为了使用从the动脉获得的脉搏波测量血压(BP),我们开发了一种带Hall装置的无袖腕式脉搏计。在分析从霍尔器件获得的非加压脉搏波的分析(检测永磁体磁场的微小变化)的基础上,提出了一种用于估计东方诊断医学中血压的新算法。使用从涉及13位受试者的临床试验中获得的脉搏波数据,我们得出了用于回归分析的方程式。该方程可用于确定相关因子。高血压的估计值和测量值的标准偏差为3.5 mmHg,低血压的标准偏差为3.8 mmHg,接近BP国际标准。