Urine analysis performed by flow cytometry: reference range determination and comparison to morphological findings, dipstick chemistry and bacterial culture results--a multicenter study.

机译:通过流式细胞仪进行尿液分析:参考范围确定以及与形态学发现,量油尺化学和细菌培养结果的比较-多中心研究。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

AIM: To validate whether quantitative flow-cytometric analysis of particulate matter in urine would allow for accurate and rapid enumeration of red blood cells (RBC), leukocytes (WBC), squamous epithelial cells (EC), casts, and bacteria, a Sysmex UF-100 analyzer was tested in a multicenter study. MATERIAL AND METHODS: At first, reference values were established and found to be < 14 for RBC, < 16 for WBC, < 9 for EC, < 2 for casts and < 173 for bacteria, respectively (counts per microl; 97.5 percentile). Due to the wide use of dipstick and microscopic sediment analysis in routine urine diagnostics, comparative studies on 950 random urine samples were performed. Bacterial counting combined with WBC enumeration was further compared in 266 routine urinary microbiologic cultures. RESULTS: Good correlations were found comparing UF-100 results of RBC (r = 0.89), WBC (r = 0.94), and EC (r = 0.74) with Fuchs-Rosenthal Chamber (FRC) counts. However, some misclassification of casts (r = 0.32) could be observed. Correlations of UF-100 with dipstick and sediment testing was significant (p < 0.001), but the scatter of the latter two methods is too wide to consider them as quantitative methods. Promising results further revealed that the analyzer has a good negative predictive value (NPV) for microbiologically negative cultures, especially for cultures with bacterial counts of 10(5)/l (NPV = 95%). CONCLUSION: The analyzer is capable of providing rapid and reliable urine analysis of cellular particles avoiding the known imprecision of dipstick and sediment methodology. Thus, when used in an algorithm, combined with dipstick or quantitative urine chemistry analysis (for hemoglobin, esterase, protein, glucose, etc.), this analyzer might serve as a rapid and accurate screening tool in routine urine analysis, thereby reducing manual reviewing rate as well as the number of missed samples, compared to screening with dipstick alone.
机译:目的:为了验证尿液中颗粒物的定量流式细胞分析是否可以准确快速地计数红细胞(RBC),白细胞(WBC),鳞状上皮细胞(EC),管型和细菌,Sysmex UF -100分析仪在多中心研究中进行了测试。材料和方法:首先,建立参考值,发现RBC的参考值小于14,WBC的参考值小于16,EC的参考值小于9,铸模的参考值小于2,细菌的参考值小于173(每微升数; 97.5个百分点)。由于在常规尿液诊断中广泛使用了量油尺和微观沉积物分析技术,因此对950个随机尿液样品进行了比较研究。在266例常规泌尿微生物培养物中进一步比较了细菌计数和WBC计数。结果:将UF-100的RBC(r = 0.89),WBC(r = 0.94)和EC(r = 0.74)与Fuchs-Rosenthal Chamber(FRC)计数进行比较,发现具有良好的相关性。但是,可以观察到铸型的一些错误分类(r = 0.32)。 UF-100与量油尺和沉淀物测试的相关性很显着(p <0.001),但是后两种方法的分散性太宽,无法将其视为定量方法。有希望的结果进一步表明,该分析仪对微生物阴性培养物具有良好的阴性预测值(NPV),尤其对于细菌计数为10(5)/ l(NPV = 95%)的培养物。结论:该分析仪能够对细胞颗粒进行快速可靠的尿液分析,从而避免了已知的不精确的量油计和沉积物方法。因此,当用于算法时,结合量油尺或定量尿液化学分析(用于血红蛋白,酯酶,蛋白质,葡萄糖等),此分析仪可作为常规尿液分析中的快速准确的筛查工具,从而减少了人工检查与仅使用量油尺进行筛查相比,检测率和漏检样品的数量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号