CT scan parameters and radiation dose: Practical advice for radiologists

机译:CT扫描参数和辐射剂量:放射科医生的实用建议

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Although there has been increasing recognition of the importance of reducing radiation dose when performing multidetector CT examinations, the increasing complexity of CT scanner technology, as well as confusion about the importance of many different CT scan parameters, has served as an impediment to radiologists seeking to create lower dose protocols. The authors seek to guide radiologists through the manipulation of 8 fundamental CT scan parameters that can be altered or optimized to reduce patient radiation dose, including detector configuration, tube current, tube potential, reconstruction algorithm, patient positioning, scan range, reconstructed slice thickness, and pitch. Although there is always an inherent trade-off between image quality or noise and patient radiation dose, in many cases, a reasoned manipulation of these 8 parameters can allow the safer imaging of patients (with lower dose) while preserving diagnostic image quality.
机译:尽管越来越多的人意识到在执行多探测器CT检查时降低辐射剂量的重要性,但是CT扫描仪技术的日益复杂以及对许多不同CT扫描参数的重要性的困惑,一直困扰着寻求放射治疗的放射科医生。创建较低剂量的方案。作者试图通过8个基本的CT扫描参数的操作来指导放射科医生,这些8个基本的CT扫描参数可以更改或优化以减少患者的辐射剂量,包括检测器配置,管电流,管电位,重建算法,患者位置,扫描范围,重建的切片厚度,和音高。尽管在图像质量或噪声与患者辐射剂量之间始终存在内在的取舍,但是在许多情况下,对这8个参数进行合理的操作可以在保留诊断图像质量的同时,对患者(剂量较低)进行更安全的成像。