MS/MS of Ions in a Low Pressure Linear Ion Trap using a Pulsed Gas

机译:使用脉冲气体的低压线性离子阱中离子的MS / MS

获取原文
获取原文并翻译 | 示例
           

摘要

A pulsed valve was used to increase the pressure within the trapping region of a low-pressure linear ion trap by situating the pulsed valve close to the ion trapping region. The pressure was estimated to increase from a background pressure of 3.5e-5 Torr of nitrogen to 0.49 mTorr at the center of the trap. The increased pressure allowed excitation periods to be reduced from 100 to 25 ms without suffering losses in MS/MS efficiency during dipolar excitation. The reduction in excitation period translates into an increase in the overall duty cycle of the scan.
机译:通过使脉冲阀靠近离子捕获区域,使用脉冲阀来增加低压线性离子阱的捕获区域内的压力。估计压力从捕集阱中心的3.5e-5 Torr的氮气本底压力增加到0.49 mTorr。增大的压力使激励周期从100毫秒减少到25毫秒,而不会在偶极激励期间降低MS / MS效率。激发周期的减少转化为扫描总占空比的增加。