首页> 外文期刊>Journal of spinal disorders & techniques. >Prospective comparison of virtual fluoroscopy to fluoroscopy and plain radiographs for placement of lumbar pedicle screws.
【24h】

Prospective comparison of virtual fluoroscopy to fluoroscopy and plain radiographs for placement of lumbar pedicle screws.

机译:虚拟透视与透视和X线平片对腰椎椎弓根螺钉放置的前瞻性比较。

获取原文
获取原文并翻译 | 示例
           

摘要

Fluoroscopy-based frameless stereotactic systems provide feedback to the surgeon using virtual fluoroscopic images. The real-life accuracy of these virtual images has not been compared with traditional fluoroscopy in a clinical setting. We prospectively studied 23 consecutive cases. In two cases, registration errors precluded the use of virtual fluoroscopy. Pedicle probes placed with virtual fluoroscopic imaging were imaged with traditional fluoroscopy in the remaining 21 cases. Position of the probes was judged to be ideal, acceptable but not ideal, or not acceptable based on the traditional fluoroscopic images. Virtual fluoroscopy was used to place probes in for 97 pedicles from L1 to the sacrum. Eighty-eight probes were judged to be in ideal position, eight were judged to be acceptable but not ideal, and one probe was judged to be in an unacceptable position. This probe was angled toward an adjacent disc space. Therefore, 96 of 97 probes placed using virtual fluoroscopy were found to be in an acceptable position. The positive predictive value for acceptable screw placement with virtual fluoroscopy compared with traditional fluoroscopy was 99%. A probe placed with virtual fluoroscopic guidance will be judged to be in an acceptable position when imaged with traditional fluoroscopy 99% of the time.
机译:基于透视的无框立体定向系统使用虚拟透视图像向外科医生提供反馈。在临床环境中,这些虚拟图像的真实准确性没有与传统的透视检查相提并论。我们前瞻性地研究了23个连续病例。在两种情况下,配准错误排除了使用虚拟透视检查的可能性。在剩余的21例病例中,使用传统的透视检查技术对放置有虚拟透视检查技术的椎弓根探头进行成像。基于传统的透视图像,探针的位置被判断为理想,可接受但不理想或不可接受。使用虚拟荧光透视法将探针放置在从L1到the骨的97个椎弓根中。八十八个探针被判定为处于理想位置,八个探针被判定为可接受但不理想,并且一个探针被判定为处于不可接受的位置。该探针朝向相邻的磁盘空间倾斜。因此,发现使用虚拟荧光检查法放置的97个探针中有96个处于可接受的位置。与传统的荧光检查相比,虚拟荧光检查对可接受的螺钉放置的阳性预测值为99%。当使用传统的透视检查技术在99%的时间对图像进行虚拟透视检查时,将其放置在可接受的位置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号