The neutron dose equivalent rate (DR) leaking from the self-shielded cyclotron was measured using an FHT-751 neutron counting system in the facility at the Cyclotron Center of Chung Shan Medical University. This system was calibrated using two Cf-252 neutron sources and simulated according to MCNP code. The results show various DRs up to 120 mu Sv h(-1) in the cyclotron room. Two-dimensional distributions of measured neutron DRs indicate an explicit, heavy leakage of neutrons through the self-shielded interlock. The neutron DR of the operating cyclotron that is hazardous to the health of medical personal and the public is evaluated.
机译:使用中山医科大学回旋加速器中心设施中的FHT-751中子计数系统测量自屏蔽回旋加速器泄漏的中子剂量当量率(DR)。该系统使用两个Cf-252中子源进行了校准,并根据MCNP代码进行了仿真。结果显示回旋加速器室内的各种DR高达120μSv h(-1)。被测中子DR的二维分布表明,中子通过自屏蔽联锁装置明显泄漏。评估正在运行的回旋加速器的中子DR对医疗人员和公众的健康有害。